Abstract
An automated analyzer for vancomycin in rat plasma by column-switching high-performance liquid chromatography (HPLC) with UV detection was developed. The method includes in-line extraction of vancomycin by ion-exchange cartridge column and a separation on a reversed-phase column with UV detection at 215 nm. Plasma samples were diluted by mobile phase solution and directly injected to HPLC. Vancomycin was quantitatively recovered from rat plasma samples. The separation was completed within 15 min. The calibration curve was linear over the range from 0.5 to 100 microg/mL with the detection and quantification limits of 0.5 microg/mL (2.5 ng on column; signal-to-noise ratio = 3). The values of precision in intra- and inter-day assays (n = 3) were less than 1.92 and 3.69%, respectively. This method does not require time-consuming pre-treatment and is suitable for the routine assay of plasma samples. 2004 John Wiley & Sons, Ltd.Entities: