Abstract
During cardiopulmonary resuscitation, oxygen extension tubing was mistakenly connected into the tracheal tube connector. As a result, the patient recovering from hypovolemic cardiac arrest died because of pulmonary barotrauma. Similar cases have been reported. We suggest that a worldwide equipment performance standard be developed to prevent future occurrences and describe an example of connectors that would prevent such misconnections.Entities:
Mesh:
Year: 2004 PMID: 15385369 DOI: 10.1213/01.ANE.0000130619.77475.8D