A block design for split-field tests of accelerator alignment.

Eric C Ford1, Wendell R Lutz.   

Abstract

We present a block design that facilitates split-field tests, a technique frequently employed to assess several potential misalignments in linear accelerators. Two tests are possible with this device using: (1) opposed collimator angles and (2) opposed gantry angles. The first test is sensitive to jaw symmetry only. In the second test, the alignment of one field edge indicates gantry sag, while the other field edge is sensitive to the combined effects of jaw asymmetry, misalignment of the beam spot, and misalignment of the gantry and collimator rotation axes. The block arrangement consists of a pattern of three polygons cast in lead or Cerrobend. The assembly mounts in the accessory tray of the accelerator and allows all necessary exposures to be taken in a single room entry. Any field size up to 24 cm can be tested with the assembly constructed here. A simultaneous test of light field-radiation field coincidence can also be accommodated. Exposed films with a prototype device show that offsets of +/-1 mm are readily visible. The block design facilitates the required measurements for split-field tests and makes this a practical technique for routine quality assurance on a linear accelerator.

Entities:  

Mesh:

Year:  2004        PMID: 15377099     DOI: 10.1118/1.1773621

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy.

Authors:  Weiliang Du; Song Gao; Xiaochun Wang; Rajat J Kudchadker
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.