| Literature DB >> 15373443 |
David S Reichmuth1, Timothy J Shepodd, Brian J Kirby.
Abstract
A high-pressure (> 3 MPa) on-chip injector has been developed for microchip applications including HPLC. The mechanical injector is implemented using in situ photopolymerization of fluorinated acrylates inside wet-etched silica microchips. The injector allows reproducible injections as small as 180 pL with < 250 ms duration. The injector operated robustly over 60 days and over 1000 injections. The injector is unique among polymer-based valves as it functions in aqueous, acetonitrile, and mixed buffers at high pressures without detectable leakage. Copyright 2004 American Chemical SocietyEntities:
Mesh:
Substances:
Year: 2004 PMID: 15373443 DOI: 10.1021/ac0493572