Ankle arthrodesis using dowel bone graft and cancellous-bone screws: a mechanical study in cadavers.

Abbas Karladani1, Georgias Digas, Lars Ekström, Janos Swanpalmer, Jorma Styf, Tommy Hansson.   

Abstract

BACKGROUND: There are numerous techniques for ankle arthrodesis.
MATERIAL AND METHODS: We studied the stability of an ankle arthrodesis with the use of two cross-screws combined with dowel graft technique in 14 fresh-frozen ankles from amputees. The arthrodesis was performed by a dowel graft technique, in the coronal plane in 7 specimens (group 1) and in the diagonal plane in the remaining 7 (group 2). Allocation into the two groups took place after normalization according to the bone mineral content of the specimens. All ankles were fixed with two 6.5-mm diameter cancellous-bone screws, with the ankle in neutral position. Tibiotalar motion was measured during 5 Nm torque loading in plantar flexion-dorsiflexion, abduction-adduction and eversion-inversion planes of the ankle joint. Two testing sequences were performed and the mean value of both tests in each direction was used for the analysis.
RESULTS: In all 3 planes, the torque applied caused more motion with the dowel graft in the coronal plane (group 1) than with the graft in the diagonal plane (group 2). The differences between the two groups were significant for abduction-adduction and eversion-inversion planes. Statistical analysis of differences between the first and second test showed significantly greater stiffness in group 2 than in group 1 in plantar flexion-dorsiflexion torque.
INTERPRETATION: The placement of a dowel graft in the diagonal plane in comparison to the coronal plane significantly increased the initial stability of an ankle arthrodesis. Copyright 2004 Taylor & Francis

Entities:  

Mesh:

Year:  2004        PMID: 15370593     DOI: 10.1080/00016470410001268-1

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  2 in total

1.  [A new technique for transcrestal sinus floor elevation with press-fit bone cylinders (dowel lift): short communication of the first in vitro results].

Authors:  G F Draenert; M Ehrenfeld; W Eisenmenger
Journal:  Mund Kiefer Gesichtschir       Date:  2007-01

2.  Small joint arthrodesis technique using a dowel bone graft in a rabbit model.

Authors:  Ki-Tae Na; Yoon-Min Lee; Jae-Hoon Choi; Seok-Whan Song
Journal:  PLoS One       Date:  2019-12-30       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.