Comparison of spot tests with AdultaCheck 6 and Intect 7 urine test strips for detecting the presence of adulterants in urine specimens.

Amitava Dasgupta1, Omar Chughtai, Christina Hannah, Bonnette Davis, Alice Wells.   

Abstract

BACKGROUND: Several adulterants are used to mask tests for abused drugs in urine. Adulterants such as "Klear" and "Whizzies" contain potassium nitrite while "Urine Luck" contains pyridinium chlorochromate (PCC). The presence of these adulterants cannot be detected by routine specimen integrity check (pH, specific gravity, creatinine and temperature). We previously reported the development of rapid spot tests to detect the presence of these adulterants. AdultaCheck 6 and Intect 7 urine test strips are commercially available for detecting the presence of these adulterants along with specific gravity, creatinine and pH in urine.
METHODS: The performance of these two test strips for detecting adulterants was compared with the results obtained by spot tests.
RESULTS: Both AdultaCheck 6 and Intect 7 effectively detected the presence of nitrite and pyridinium chlorochromate in urine. Moreover, both test strips successfully detected the presence of glutaraldehyde, for which no spot test is currently available. High amount of glucose and ascorbic acid did not cause any false positive result with AdultaCheck 6 or Intect 7.
CONCLUSIONS: Both AdultaCheck 6 and Intect 7 can be used for checking the integrity of a urine specimen submitted for drugs of abuse testing.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15369731     DOI: 10.1016/j.cccn.2004.01.003

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  A case of psychosis after use of a detoxification kit and a review of techniques, risks, and regulations associated with the subversion of urine drug tests.

Authors:  Moneeshindra Singh Mittal; Rachna Kalia; Ahsan Y Khan
Journal:  Prim Care Companion CNS Disord       Date:  2011

2.  Rational Urine Drug Monitoring in Patients Receiving Opioids for Chronic Pain: Consensus Recommendations.

Authors:  Charles E Argoff; Daniel P Alford; Jeffrey Fudin; Jeremy A Adler; Matthew J Bair; Richard C Dart; Roy Gandolfi; Bill H McCarberg; Steven P Stanos; Jeffrey A Gudin; Rosemary C Polomano; Lynn R Webster
Journal:  Pain Med       Date:  2018-01-01       Impact factor: 3.750

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.