Literature DB >> 15367767

In vivo dosimetry for external photon treatments of head and neck cancers by diodes and TLDS.

C J Tung1, H C Wang, S H Lo, J M Wu, C J Wang.   

Abstract

In vivo dosimetry was implemented for treatments of head and neck cancers in the large fields. Diode and thermoluminescence dosemeter (TLD) measurements were carried out for the linear accelerators of 6 MV photon beams. ESTRO in vivo dosimetry protocols were followed in the determination of midline doses from measurements of entrance and exit doses. Of the fields monitored by diodes, the maximum absolute deviation of measured midline doses from planned target doses was 8%, with the mean value and the standard deviation of -1.0 and 2.7%. If planned target doses were calculated using radiological water equivalent thicknesses rather than patient geometric thicknesses, the maximum absolute deviation dropped to 4%, with the mean and the standard deviation of 0.7 and 1.8%. For in vivo dosimetry monitored by TLDs, the shift in mean dose remained small but the statistical precision became poor.

Entities:  

Mesh:

Year:  2004        PMID: 15367767     DOI: 10.1093/rpd/nch358

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Evaluation of off-axis wedge correction factor using diode dosimeters for estimation of delivered dose in external radiotherapy.

Authors:  Mahmoud Allahverdi; Alireza Mohammadkarim; Mahbod Esfehani; Hasanali Nedaie; Alireza Shirazi; Ghazale Geraily
Journal:  J Med Phys       Date:  2012-01

2.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

3.  Evaluation of target dose based on water-equivalent thickness in external beam radiotherapy.

Authors:  Behnaz Ghanbar Moghaddam; Masoud Vahabi-Moghaddam; Alireza Sadremomtaz
Journal:  J Med Phys       Date:  2013-01

4.  Measurement comparison and Monte Carlo analysis for volumetric-modulated arc therapy (VMAT) delivery verification using the ArcCHECK dosimetry system.

Authors:  Mu-Han Lin; Sion Koren; Iavor Veltchev; Jinsheng Li; Lu Wang; Robert A Price; C-M Ma
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

  4 in total