An evaluation of the Brüel and Kjaer monitor 1304.

H B McPeak1, E Palayiwa, G C Robinson, M K Sykes.   

Abstract

A laboratory evaluation was performed on the Brüel and Kjaer multigas monitor 1304, incorporating a pulse oximeter. The instrument was tested for accuracy, stability, response and delay times, frequency response and the effects of water vapour, alcohol, cyclopropane and sevoflurane. The instrument's performance was found to be within or very close to the manufacturer's specifications for accuracy, stability and response and delay times. It was unaffected by water vapour and alcohol and the effect of cyclopropane on the vapour channel was lower than has been reported for other analysers. The response to sevoflurane was of the same order as that of the other vapours. A 90% response to square wave changes of gas composition was maintained up to 60 breaths.min-1 for CO2, O2, and N2O and up to 40 breaths.min-1 for the vapours when the nafion sampling tube was used.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1536404     DOI: 10.1111/j.1365-2044.1992.tb01953.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

1.  Effects of nitrous oxide on psychological performance. A dose-response study using inhalation of concentrations up to 15%.

Authors:  P J Armstrong; C Morton; W Sinclair; B Tiplady
Journal:  Psychopharmacology (Berl)       Date:  1995-02       Impact factor: 4.530

2.  Carbon dioxide analysers: accuracy, alarm limits and effects of interfering gases.

Authors:  R Lauber; B Seeberger; A M Zbinden
Journal:  Can J Anaesth       Date:  1995-07       Impact factor: 5.063

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.