A dual closed-loop control system for mechanical ventilation.

Fleur Tehrani1, Mark Rogers, Takkin Lo, Thomas Malinowski, Samuel Afuwape, Michael Lum, Brett Grundl, Michael Terry.   

Abstract

OBJECTIVE: Closed-loop mechanical ventilation has the potential to provide more effective ventilatory support to patients with less complexity than conventional ventilation. The purpose of this study was to investigate the effectiveness of an automatic technique for mechanical ventilation.
METHODS: Two closed-loop control systems for mechanical ventilation are combined in this study. In one of the control systems several physiological data are used to automatically adjust the frequency and tidal volume of breaths of a patient. This method, which is patented under US Patent number 4986268, uses the criterion of minimal respiratory work rate to provide the patient with a natural pattern of breathing. The inputs to the system include data representing CO2 and O2 levels of the patient as well as respiratory compliance and airway resistance. The I:E ratio is adjusted on the basis of the respiratory time constant to allow for effective emptying of the lungs in expiration and to avoid intrinsic positive end expiratory pressure (PEEP). This system is combined with another closed-loop control system for automatic adjustment of the inspired fraction of oxygen of the patient. This controller uses the feedback of arterial oxygen saturation of the patient and combines a rapid stepwise control procedure with a proportional-integral-derivative (PID) control algorithm to automatically adjust the oxygen concentration in the patient's inspired gas. The dual closed-loop control system has been examined by using mechanical lung studies, computer simulations and animal experiments.
RESULTS: In the mechanical lung studies, the ventilation controller adjusted the breathing frequency and tidal volume in a clinically appropriate manner in response to changes in respiratory mechanics. The results of computer simulations and animal studies under induced disturbances showed that blood gases were returned to the normal physiologic range in less than 25 s by the control system. In the animal experiments understeady-state conditions, the maximum standard deviations of arterial oxygen saturation and the end-tidal partial pressure of CO2 were +/- 1.76% and +/- 1.78 mmHg, respectively.
CONCLUSION: The controller maintained the arterial blood gases within normal limits under steady-state conditions and the transient response of the system was robust under various disturbances. The results of the study have showed that the proposed dual closed-loop technique has effectively controlled mechanical ventilation under different test conditions.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15362273     DOI: 10.1023/b:jocm.0000032744.99885.38

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  22 in total

1.  A servomechanism for automatic regulation of pulmonary ventilation.

Authors:  G A SAXTON; G MYERS
Journal:  J Appl Physiol       Date:  1957-09       Impact factor: 3.531

2.  Mechanics of breathing in man.

Authors:  A B OTIS; W O FENN; H RAHN
Journal:  J Appl Physiol       Date:  1950-05       Impact factor: 3.531

3.  Closed-loop control of an anesthesia delivery system: development and animal testing.

Authors:  R G Ritchie; E A Ernst; B L Pate; J D Pearson; L C Sheppard
Journal:  IEEE Trans Biomed Eng       Date:  1987-06       Impact factor: 4.538

4.  New technique for servo-control of arterial oxygen tension in preterm infants.

Authors:  I R Beddis; P Collins; N M Levy; S Godfrey; M Silverman
Journal:  Arch Dis Child       Date:  1979-04       Impact factor: 3.791

5.  A feedback controller for supplemental oxygen treatment of newborn infants: a simulation study.

Authors:  F T Tehrani; A R Bazar
Journal:  Med Eng Phys       Date:  1994-07       Impact factor: 2.242

6.  A mathematical model of the human respiratory system.

Authors:  W F Fincham; F T Tehrani
Journal:  J Biomed Eng       Date:  1983-04

7.  Carbon dioxide and oxygen blood levels with a carbon dioxide controlled artificial respirator.

Authors:  M J FRUMIN; N A BERGMAN; D A HOLADAY
Journal:  Anesthesiology       Date:  1959 May-Jun       Impact factor: 7.892

8.  Closed-loop control if the inspired fraction of oxygen in mechanical ventilation.

Authors:  Fleur Tehrani; Mark Rogers; Takkin Lo; Thomas Malinowski; Samuel Afuwape; Michael Lum; Brett Grundl; Michael Terry
Journal:  J Clin Monit Comput       Date:  2002-08       Impact factor: 2.502

9.  A microprocessor based feedback controller for mechanical ventilation.

Authors:  K B Ohlson; D R Westenskow; W S Jordan
Journal:  Ann Biomed Eng       Date:  1982       Impact factor: 3.934

10.  A feedback controller for ventilatory therapy.

Authors:  F W Chapman; J C Newell; R J Roy
Journal:  Ann Biomed Eng       Date:  1985       Impact factor: 3.934

View more
  13 in total

1.  A model-based decision support system for critiquing mechanical ventilation treatments.

Authors:  Fleur T Tehrani; Soraya Abbasi
Journal:  J Clin Monit Comput       Date:  2012-04-25       Impact factor: 2.502

2.  Evaluation of a computerized system for mechanical ventilation of infants.

Authors:  Fleur T Tehrani; Soraya Abbasi
Journal:  J Clin Monit Comput       Date:  2009-03-05       Impact factor: 2.502

Review 3.  Automatic control of mechanical ventilation. Part 1: theory and history of the technology.

Authors:  Fleur T Tehrani
Journal:  J Clin Monit Comput       Date:  2008-11-16       Impact factor: 2.502

4.  A knowledge- and model-based system for automated weaning from mechanical ventilation: technical description and first clinical application.

Authors:  Dirk Schädler; Stefan Mersmann; Inéz Frerichs; Gunnar Elke; Thomas Semmel-Griebeler; Oliver Noll; Sven Pulletz; Günther Zick; Matthias David; Wolfgang Heinrichs; Jens Scholz; Norbert Weiler
Journal:  J Clin Monit Comput       Date:  2013-07-28       Impact factor: 2.502

5.  Adaptive support ventilation is a patented technology.

Authors:  Fleur T Tehrani
Journal:  Intensive Care Med       Date:  2008-09-25       Impact factor: 17.440

6.  Automatic Detection of Endotracheal Intubation During the Anesthesia Procedure.

Authors:  Ali Jalali; Mohamed Rehman; Arul Lingappan; C Nataraj
Journal:  J Dyn Syst Meas Control       Date:  2016-08-09       Impact factor: 1.372

Review 7.  Machine Learning and Artificial Intelligence in Neurocritical Care: a Specialty-Wide Disruptive Transformation or a Strategy for Success.

Authors:  Fawaz Al-Mufti; Michael Kim; Vincent Dodson; Tolga Sursal; Christian Bowers; Chad Cole; Corey Scurlock; Christian Becker; Chirag Gandhi; Stephan A Mayer
Journal:  Curr Neurol Neurosci Rep       Date:  2019-11-13       Impact factor: 5.081

8.  Flex: a new computerized system for mechanical ventilation.

Authors:  Fleur T Tehrani; James H Roum
Journal:  J Clin Monit Comput       Date:  2008-03-07       Impact factor: 2.502

Review 9.  Automatic control of mechanical ventilation. Part 2: the existing techniques and future trends.

Authors:  Fleur T Tehrani
Journal:  J Clin Monit Comput       Date:  2008-11-20       Impact factor: 2.502

Review 10.  Newer nonconventional modes of mechanical ventilation.

Authors:  Preet Mohinder Singh; Anuradha Borle; Anjan Trikha
Journal:  J Emerg Trauma Shock       Date:  2014-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.