Characterisation of neutron fields around high-energy x-ray radiotherapy machines.

M Králík1, K Turek.   

Abstract

Photoneutron spectra around the treatment bed of a Varian Clinac 2100C machine were measured using a Bonner sphere spectrometer. To overcome problems with pulse pile-up and detection of non-neutron-induced events, the active detector of thermal neutrons normally used at the centre of the spheres was replaced by a sandwich of four CR-39 track detectors interleaved with 10B radiators. Track densities measured for the CR-39 detectors in Bonner spheres were used for the unfolding of neutron spectra. Neutron fluence and ambient dose equivalent for the whole energy range and partial energy intervals were derived from the neutron spectra.

Entities:  

Mesh:

Year:  2004        PMID: 15353699     DOI: 10.1093/rpd/nch274

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  5 in total

Review 1.  A review on photoneutrons characteristics in radiation therapy with high-energy photon beams.

Authors:  Alireza Naseri; Asghar Mesbahi
Journal:  Rep Pract Oncol Radiother       Date:  2010-09-22

2.  Secondary radiation dose during high-energy total body irradiation.

Authors:  M Janiszewska; K Polaczek-Grelik; M Raczkowski; B Szafron; A Konefał; W Zipper
Journal:  Strahlenther Onkol       Date:  2014-03-06       Impact factor: 3.621

3.  Cephalometric assessment of the axial inclination of upper and lower incisors in relation to the third-order angle.

Authors:  Michael Knösel; Rengin Attin; Dietmar Kubein-Meesenburg; Reza Sadat-Khonsari
Journal:  J Orofac Orthop       Date:  2007-05       Impact factor: 1.938

4.  A neutron track etch detector for electron linear accelerators in radiotherapy.

Authors:  Branko Vukovic; Dario Faj; Marina Poje; Maja Varga; Vanja Radolic; Igor Miklavcic; Ana Ivkovic; Josip Planinic
Journal:  Radiol Oncol       Date:  2010-03-18       Impact factor: 2.991

5.  Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations.

Authors:  Hassan Ali Nedaie; Hoda Darestani; Nooshin Banaee; Negin Shagholi; Kheirollah Mohammadi; Arjang Shahvar; Esmaeel Bayat
Journal:  J Med Phys       Date:  2014-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.