Literature DB >> 15353016

Reliability of the Prognos electrodermal device for measurements of electrical skin resistance at acupuncture points.

Agatha P Colbert1, Richard Hammerschlag, Mikel Aickin, James McNames.   

Abstract

OBJECTIVES: (1) To characterize and calibrate an electrodermal screening device, Prognos. (2) To replicate a previous test-retest reliability study of this device with measurements of electrical skin resistance (ESR) at 24 Jing-well acupuncture points (APs). (3) To determine measurement precision in three successively more exacting trial protocols on the same set of subjects. SETTINGS: Oregon College of Oriental Medicine and Portland State University, Portland, OR. INSTRUMENTS: The Prognos device was electrically characterized by a team of research engineers at the Biomedical Signal Processing Laboratory of Portland State University. They determined that Prognos measures the average direct-current (DC) resistance between a metallic wrist strap and an electrode probe tip. The probe tip is connected to a linear spring set to trigger with an optically generated signal at a deflection of 2.62 mm, which corresponds to an average applied force of 2.68 +/- 0.04 N (mean +/- standard deviation [SD], n = 6). They also determined that the device quantifies resistance by applying a 1.1 microA current for an average of 223 +/- 3 ms (n = 7). When calibrated against a series of known resistors, Prognos measures accurately in the range of 150 kOmega to 14.3 MOmega with an error of less than 0.4%.
SUBJECTS: Thirty-one (31) healthy volunteers, 17 females and 14 males, 23-63 years of age. RESULTS OF RELIABILITY TEST-RETEST: The mean reliability of a single measurement was; 0.758 for a standard measurement protocol of four sequential sweeps of 24 Jing-well (Ting) APs; 0.851 for four sequential sweeps after ink-marking the APs; and 0.961 for four rapid repeat measurements at each inked AP. Mean absolute values of ESR decreased between the standard and marked protocols, but not between the marked and rapid repeat protocols.
CONCLUSIONS: Prognos performs accurately, against known resistors over the reported range of ESR. The reliability in the standard protocol (r = 0.758) is comparable to the reliability of 0.721 demonstrated under similar conditions by other investigators. Marking APs, and performing measurements in a rapid sequence, increases reliability of ESR measurements. Increased reliability in the second and third protocols is associated with decreased mean ESR values which may be related to increased accuracy of Prognos probe placement and/or inking the APs.

Mesh:

Year:  2004        PMID: 15353016     DOI: 10.1089/acm.2004.10.610

Source DB:  PubMed          Journal:  J Altern Complement Med        ISSN: 1075-5535            Impact factor:   2.579


  11 in total

1.  Development and evaluation of a novel four-electrode device system for monitoring skin impedance.

Authors:  Rezaei Shima; Zhu Jiang; She Yan Fen; Afzal-Aghayi Monnavar; Khorsand Ali
Journal:  Afr J Tradit Complement Altern Med       Date:  2012-07-01

2.  Reliability of AcuGraph system for measuring skin conductance at acupoints.

Authors:  Scott D Mist; Mikel Aickin; Paul Kalnins; Jim Cleaver; Roger Batchelor; Tracy Thorne; Steve Chamberlin; Kim Tippens; Agatha P Colbert
Journal:  Acupunct Med       Date:  2011-05-20       Impact factor: 2.267

3.  Attitudes and interests toward research among students at two colleges of acupuncture and Oriental medicine.

Authors:  Peter M Wayne; Richard Hammerschlag; Jacqueline Savetsky-German; Tim F Chapman
Journal:  Explore (NY)       Date:  2010 Jan-Feb       Impact factor: 1.775

4.  A multichannel system for continuous measurements of skin resistance and capacitance at acupuncture points.

Authors:  Agatha P Colbert; Adrian Larsen; Steve Chamberlin; Carrie Decker; Heather C Schiffke; William L Gregory; Tran Thong
Journal:  J Acupunct Meridian Stud       Date:  2009-12

5.  The Study of Dynamic Characteristic of Acupoints Based on the Primary Dysmenorrhea Patients with the Tenderness Reflection on Diji (SP 8).

Authors:  Sheng Chen; Yanhuan Miao; Yinan Nan; Yanping Wang; Qi Zhao; Enhui He; Yini Sun; Jiping Zhao
Journal:  Evid Based Complement Alternat Med       Date:  2015-06-16       Impact factor: 2.629

6.  Evaluation of Applied Kinesiology meridian techniques by means of surface electromyography (sEMG): demonstration of the regulatory influence of antique acupuncture points.

Authors:  Roy Moncayo; Helga Moncayo
Journal:  Chin Med       Date:  2009-05-29       Impact factor: 5.455

7.  Measurement reliability for acupoint activity determined with the prognos ohmmeter.

Authors:  L Turner; W Linden; A Talbot Ellis; R Millman
Journal:  Appl Psychophysiol Biofeedback       Date:  2010-09

8.  Skin impedance measurements for acupuncture research: development of a continuous recording system.

Authors:  Agatha P Colbert; Jinkook Yun; Adrian Larsen; Tracy Edinger; William L Gregory; Tran Thong
Journal:  Evid Based Complement Alternat Med       Date:  2007-06-15       Impact factor: 2.629

9.  Altered Impedance of Ear Acupuncture Point MT2 in Breast Cancer Patients: A Preliminary Observation.

Authors:  Yine Hu; Huayuan Yang; Pin Wang; Tangyi Liu; Wenchao Tang
Journal:  Evid Based Complement Alternat Med       Date:  2015-10-04       Impact factor: 2.629

10.  Imaging-guided hyperstimulation analgesia in low back pain.

Authors:  Miguel Gorenberg; Kobi Schwartz
Journal:  J Pain Res       Date:  2013-06-25       Impact factor: 3.133

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.