Abstract
Recent developments in new electrolytic devices that utilize the electrolysis of water and charge-selective electromigration of ions through ion-exchange media have significantly changed the routine operation of ion chromatographic methods. Examples of these new electrolytic devices include on-line eluent generators that produce high-purity electrolyte eluents using deionized water as the carrier stream, continuously regenerated trap columns that remove ionic contaminants in the eluents, and continuously regenerated suppressors that reduce eluent background conductance prior to conductivity detection. The combined use of these electrolytic devices has made it possible to perform various ion chromatographic separations using only deionized water as the mobile phase. This paper reviews the operation principles of these electrolytic devices and their applications in the ion chromatographic determination of anionic and cationic analytes.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15345293 DOI: 10.1016/j.jbbm.2004.01.004