Management of an unusual complication during placement of a pulmonary artery catheter.

Parneeta Bhatia1, Nahel N Saied, Mark E Comunale.   

Abstract

We describe an unusual complication during flotation of a pulmonary artery catheter through a preexisting percutaneous introducer sheath. A malfunctioning pulmonary artery catheter, which was placed through an introducer sheath in the right internal jugular vein, was removed. Attempts at repositioning a second pulmonary artery catheter met with resistance, and we were unable to either advance or withdraw it. Chest radiograph showed a bent introducer sheath going from the right internal jugular vein into the right subclavian vein and a pulmonary artery catheter loop. Under continuous fluoroscopy, the introducer sheath and the pulmonary artery catheter were withdrawn as one unit, which resulted in relaxation of the acute angulation in the introducer sheath and allowed the pulmonary artery catheter to unfold, thus facilitating their complete extraction. We conclude that complications may occur during placement of a pulmonary artery catheter through a well positioned introducer sheath and that fluoroscopy is a valuable tool for safe management of such a complication.

Mesh:

Year:  2004        PMID: 15333390     DOI: 10.1213/01.ANE.0000133244.59867.6F

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  5 in total

1.  Pulmonary artery catheter coiled in the main pulmonary artery trunk.

Authors:  Nidhi Aggarwal; Yizhak Kupfer; Taek Sang Yoon; Sidney Tessler
Journal:  BMJ Case Rep       Date:  2013-07-04

2.  Endovascular treatment of PA pseudoaneurysm caused by Swan-Ganz catheter.

Authors:  Piotr N Rudziński; Marcin Demkow; Ilona Michałowska; Elżbieta Abramczuk; Piotr Szymański
Journal:  Postepy Kardiol Interwencyjnej       Date:  2014-03-23       Impact factor: 1.426

3.  Pulmonary artery rupture as a complication of Swan-Ganz catheter application. Diagnosis and endovascular treatment: a single centre's experience.

Authors:  Piotr N Rudziński; Jan Henzel; Zofia Dzielińska; Barbara M Lubiszewska; Ilona Michałowska; Piotr Szymański; Radosław Pracoń; Tomasz Hryniewiecki; Marcin Demkow
Journal:  Postepy Kardiol Interwencyjnej       Date:  2016-05-11       Impact factor: 1.426

4.  Malposition of Central Venous Catheter: Presentation and Management.

Authors:  Lin Wang; Zhang-Suo Liu; Chang-An Wang
Journal:  Chin Med J (Engl)       Date:  2016-01-20       Impact factor: 2.628

5.  Cephalad misplacement of a pulmonary artery catheter in a patient with a preexisting Hickman catheter.