| Literature DB >> 15329863 |
I Losito1, L Monaci, F Palmisano, G Tantillo.
Abstract
A method based on liquid chromatography with electrospray ionisation ion trap mass spectrometry, for the determination of ochratoxin A (OTA) in meat products using ochratoxin B (OTB) as an internal standard, is described. Fragmentation patterns of OTA and OTB were studied by sequential mass spectrometry. Trace determination was then accomplished by consecutive reaction monitoring (CRM) of a fragment obtained by MS(3) experiments. This led to a better signal-to-noise ratio and to a higher specificity of the technique. The response to OTA was linear over at least one concentration decade with a limit of detection of 0.6 ng/g. The method was applied to pig tissue samples naturally contaminated by OTA. Copyright 2004 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15329863 DOI: 10.1002/rcm.1577
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419