OBJECTIVE: To compare the accuracy and clinical usefulness of the near-patient testing CoaguChek S INR monitor in rural medical practice. DESIGN, SETTING AND MAIN OUTCOME MEASURES: General practices were identified through Australian university departments of rural health. Study investigators trained general practitioners and/or practice nurses in the use of the CoaguChek S INR monitor. General practices obtained a fingerprick sample for testing with the INR monitor to compare with conventional pathology testing for accuracy. An evaluation questionnaire was administered to users of the machine to assess ease of use and clinical usefulness.
RESULTS: A total of 169 patients from 15 general practice sites provided 401 paired (CoaguChek S and laboratory) INR results. The CoaguChek S was found to be accurate when compared to laboratory INR (r = 0.89), despite complicating variables such as multiple users of the monitor and multiple laboratories used for comparison with the CoaguChek S INR. Overall, 88% of dual INR measurements were within 0.5 INR units of each other. For laboratory INR </= 1.9, 2.0-3.5 and >/= 3.6, 97%, 90% and 57% of readings were within 0.5 INR units, respectively. Clinical agreement occurred 93% and 90% of the time against published expanded and narrow criteria, respectively.
CONCLUSIONS: The routine use of near-patient testing, with appropriate training and quality assurance programs, has the potential to increase the safety and efficacy of warfarin therapy in rural and remote communities.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15315540     DOI: 10.1111/j.1440-1854.2004.00585.x

Source DB:  PubMed          Journal:  Aust J Rural Health        ISSN: 1038-5282            Impact factor:   1.662


  3 in total

1.  Accuracy and clinical utility of the CoaguChek XS portable international normalised ratio monitor in a pilot study of warfarin home-monitoring.

Authors:  Luke R Bereznicki; Shane L Jackson; Gregory M Peterson; Ella C Jeffrey; Katherine A Marsden; David M Jupe
Journal:  J Clin Pathol       Date:  2006-06-02       Impact factor: 3.411

2.  Pharmacotherapy for atrial fibrillation in elderly hospitalized patients with comorbid congestive heart failure in australia: A retrospective study.

Authors:  Lexin Wang; Shane Curran; Patrick Ball; Fiona White
Journal:  Curr Ther Res Clin Exp       Date:  2008-12

3.  A pragmatic cluster randomised controlled trial to evaluate the safety, clinical effectiveness, cost effectiveness and satisfaction with point of care testing in a general practice setting - rationale, design and baseline characteristics.

Authors:  Caroline Laurence; Angela Gialamas; Lisa Yelland; Tanya Bubner; Philip Ryan; Kristyn Willson; Briony Glastonbury; Janice Gill; Mark Shephard; Justin Beilby
Journal:  Trials       Date:  2008-08-06       Impact factor: 2.279

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.