Literature DB >> 15311836

Design of coils for millimeter- and submillimeter-sized biotelemetry.

Sérgio Francisco Pichorim1, Paulo José Abatti.   

Abstract

Biotelemetric systems, especially those that employ implanted devices, work with inductive links, where usually large circular external coils are separated by relatively large distances (dimensions of centimeters) from the small (millimeter- or submillimeter-size) implanted coils. This paper shows that, under these conditions, a simplified method for calculation of the mutual inductance (M) between the coils, avoiding elliptic integrals, can be obtained. A procedure for coil design, with maximum M between them, is also described.

Mesh:

Year:  2004        PMID: 15311836     DOI: 10.1109/TBME.2004.827542

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538