Abstract
The characteristics boron-dose enhancer (BDE) was evaluated as to the dependence on the (10)B concentration for BNCT using near-threshold (7)Li(p,n)(7)Be direct neutrons. The treatable protocol depth (TPD) was utilized as an evaluation index. MCNP-4B calculations were performed for near-threshold (7)Li(p,n)(7)Be at a proton energy of 1.900MeV and for a polyethylene BDE. Consequently, the TPD was increased by increasing T/N ratio, i.e., the ratio of the (10)B concentration in the tumor ((10)B(Tumor)) to that in the normal tissue ((10)B(Normal)), and by increasing (10)B(Tumor) and (10)B(Normal) for constant T/N ratio. It has been found that the BDE becomes unnecessary from the viewpoint of increasing the TPD, when (10)B(Tumor) is over a certain level.Entities: