Literature DB >> 15308140

In-phantom imaging of all dose components in boron neutron capture therapy by means of gel dosimeters.

G Gambarini1, V Colli, S Gay, C Petrovich, L Pirola, G Rosi.   

Abstract

The experimental method for in-phantom imaging and profiling the absorbed dose in neutron capture therapy has been improved. The method separates the contributions of the various secondary radiation components and is based on suitably designed gel dosimeters in the form of layers. The discrimination of the dose components is achieved by means of pixel-to-pixel manipulations of images obtained with gel dosimeters having different isotopic composition. Large dose images are obtainable with this method, because the layer geometry of dosimeters avoids sensible variation of neutron transport due to the isotopic composition of gel. Operation modalities aimed at attaining more reliable results have been studied. Some results, together with the results of punctual measurements performed with conventional dosimeters and with MC calculations, are here reported.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15308140     DOI: 10.1016/j.apradiso.2004.05.061

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  5 in total

Review 1.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

2.  Determining a methodology of dosimetric quality assurance for commercially available accelerator-based boron neutron capture therapy system.

Authors:  Katsumi Hirose; Takahiro Kato; Takaomi Harada; Tomoaki Motoyanagi; Hiroki Tanaka; Akihiko Takeuchi; Ryohei Kato; Shinya Komori; Yuhei Yamazaki; Kazuhiro Arai; Noriyuki Kadoya; Mariko Sato; Yoshihiro Takai
Journal:  J Radiat Res       Date:  2022-07-19       Impact factor: 2.438

3.  A novel method of estimating dose responses for polymer gels using texture analysis of scanning electron microscopy images.

Authors:  Cheng-Ting Shih; Jui-Ting Hsu; Rou-Ping Han; Bor-Tsung Hsieh; Shu-Jun Chang; Jay Wu
Journal:  PLoS One       Date:  2013-07-02       Impact factor: 3.240

4.  MAGIC polymer gel for dosimetric verification in boron neutron capture therapy.

Authors:  Jouni Uusi-Simola; Sami Heikkinen; Petri Kotiluoto; Tom Serén; Tiina Seppälä; Iiro Auterinen; Sauli Savolainen
Journal:  J Appl Clin Med Phys       Date:  2007-04-30       Impact factor: 2.102

5.  Fricke gel dosimeter with improved sensitivity for low-dose-level measurements.

Authors:  Mauro Valente; Wladimir Molina; Lila Carrizales Silva; Rodolfo Figueroa; Francisco Malano; Pedro Pérez; Mauricio Santibañez; José Vedelago
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  5 in total