Alumina hip joints characterized by run-in wear and steady-state wear to 14 million cycles in hip-simulator model.

Hironobu Oonishi1, Ian C Clarke, Victoria Good, Hirokazu Amino, Masuru Ueno.   

Abstract

The biphasic wear performance (run-in; steady-state phase) of 28-mm alumina-alumina hip implants was studied by hip simulator methods using bovine serum as the lubricant. The Biolox implants were run to 5.7 million cycles and Bioceram implants to 14.4 million cycles (Mc). Wear with all-alumina total hip replacements (THR) first showed a high wear rate of the order 1.2 mm3/Mc, lasting approximately 0.17 Mc. Overall to 0.7 Mc, the run-in phase appeared curvilinear but could be described by a linear phase averaging 0.3 mm3/Mc. From 0.7 to 1 Mc duration, the wear trend transitioned into a steady-state phase. Wear rates from 1 to 14 Mc were of the order 0.02 mm3/Mc. Surface contamination from the serum lubricant resulted in cyclic weight fluctuations of the order 0.2 mg. The transition from average run-in to steady-state phase represented a wear reduction of 13-fold. Comparing the steady-state wear value to that in standard 28-mm UHMWPE CUPS approaching 75 mm3/year, there was clearly a three-orders-of-magnitude wear superiority in favor of ceramic cups.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15307156     DOI: 10.1002/jbm.a.30021

Source DB:  PubMed          Journal:  J Biomed Mater Res A        ISSN: 1549-3296            Impact factor:   4.396


  7 in total

Review 1.  New alternate bearing surfaces in total hip arthroplasty: A review of the current literature.

Authors:  Preston W Grieco; Scott Pascal; Jared M Newman; Neil V Shah; Sarah G Stroud; Neil P Sheth; Aditya V Maheshwari
Journal:  J Clin Orthop Trauma       Date:  2017-10-27

2.  Cartilage-mimicking, high-density brush structure improves wear resistance of crosslinked polyethylene: a pilot study.

Authors:  Masayuki Kyomoto; Toru Moro; Yoshio Takatori; Hiroshi Kawaguchi; Kazuhiko Ishihara
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

3.  Natural history of squeaking after total hip arthroplasty.

Authors:  Camilo Restrepo; Wadih Y Matar; Javad Parvizi; Richard H Rothman; William J Hozack
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

Review 4.  Bioceramics for Hip Joints: The Physical Chemistry Viewpoint.

Authors:  Giuseppe Pezzotti
Journal:  Materials (Basel)       Date:  2014-06-11       Impact factor: 3.623

5.  Tribology of total hip arthroplasty prostheses: What an orthopaedic surgeon should know.

Authors:  Claude B Rieker
Journal:  EFORT Open Rev       Date:  2017-03-13

6.  No benefit on functional outcomes and dislocation rates by increasing head size to 36 mm in ceramic-on-ceramic total hip arthroplasty.

Authors:  Yu-Der Lu; Shih-Hsiang Yen; Feng-Chih Kuo; Jun-Wen Wang; Ching-Jen Wang
Journal:  Biomed J       Date:  2016-03-10       Impact factor: 4.910

7.  Effects of Surface Modification and Bulk Geometry on the Biotribological Behavior of Cross-Linked Polyethylene: Wear Testing and Finite Element Analysis.

Authors:  Kenichi Watanabe; Masayuki Kyomoto; Kenichi Saiga; Shuji Taketomi; Hiroshi Inui; Yuho Kadono; Yoshio Takatori; Sakae Tanaka; Kazuhiko Ishihara; Toru Moro
Journal:  Biomed Res Int       Date:  2015-10-25       Impact factor: 3.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.