Attenuation of intracavitary applicators in 192Ir-HDR brachytherapy.

Sung-Joon Ye1, Ivan A Brezovich, Sui Shen, Jun Duan, Richard A Popple, Prem N Pareek.   

Abstract

Unlike the penetrating monoenergetic 662 keV gamma rays emitted by 137Cs LDR sources, the spectrum of 192Ir used in HDR brachytherapy contains low-energy components. Since these are selectively absorbed by the high-atomic number materials of which intracavitary applicators are made, the traditional neglect of applicator attenuation can lead to appreciable dose errors. We investigated the attenuation effects of a uterine applicator, and of a set of commonly used vaginal cylinders. The uterine applicator consists of a stainless steel source guide tube with a wall thickness of 0.5 mm and a density of 8.02 g/cm3, whereas the vaginal cylinders consist of the same stainless steel tube plus concentric polysulfone cylinders with a radius of 1 or 2 cm and a density of 1.40 g/cm3. Monte Carlo simulations were performed to compute dose distributions for a bare 192Ir-HDR source, and for the same source located within the applicators. Relative measurements of applicator attenuation using ion-chambers (0.125 cm3) confirmed the Monte Carlo results within 0.5%. We found that the neglect of the applicator attenuation overestimates the dose along the transverse plane by up to 3.5%. At oblique angles, the longer photon path within applicators worsens the error. We defined attenuation-corrected radial dose and anisotropy functions, and applied them to a treatment having multiple dwell positions inside a vaginal cylinder.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15305463     DOI: 10.1118/1.1762791

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Dosimetric characteristics of the ¹⁹²Ir high-dose-rate afterloading brachytherapy source.

Authors:  Mahdi Sadeghi; Fatemah Taghdiri; Pooneh Saidi
Journal:  Jpn J Radiol       Date:  2011-06-30       Impact factor: 2.374

Review 2.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17

3.  Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix.

Authors:  Satish C Uniyal; Umesh C Naithani; Sunil D Sharma; Anoop K Srivastava
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

4.  Dose perturbation due to the polysulfone cap surrounding a Fletcher-Williamson colpostat.

Authors:  Michael J Price; Stephen F Kry; Patricia J Eifel; Mohammad Salehpour; Firas Mourtada
Journal:  J Appl Clin Med Phys       Date:  2010-01-28       Impact factor: 2.102

5.  Experimental verification of Advanced Collapsed-cone Engine for use with a multichannel vaginal cylinder applicator.

Authors:  Brie Cawston-Grant; Hali Morrison; Geetha Menon; Ron S Sloboda
Journal:  J Appl Clin Med Phys       Date:  2017-03-20       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.