Packed column supercritical fluid chromatography of a peroxysome proliferator-activating receptor agonist drug. Achiral and chiral purity of substance, formulation assay and its enantiomeric purity.

Olle Gyllenhaal1.   

Abstract

This paper describes packed column supercritical fluid chromatography (SFC) for the analysis of a peroxysome proliferator-activating gamma-receptor agonist that is a carboxylic acid. Evaluation of conditions for the separation of this candidate drug and related compounds in bulk substance is described. A Chiralcel OD column was used for this purpose due to its high selective retention of related substances and relative inertness, though the enantioselectivity was negligible, with methanol as polar modifier. A high enantioselectivity was obtained on Chiralpak AD and it was possible to determine the enantiomeric purity within 10 min on a 5 cm short column. Both the achiral and the chiral systems were run without acid additive in the mobile phase and the level of detection of impurities by area was about 0.1%. For the analysis of samples dissolved in water, without any isolation step, 2-propanol was used as modifier. Due to the column surface activity, evidently generated by injected water, citric acid 1 mM was included as additive in the 2-propanol in order to maintain symmetric and undistorted peak shape. The detection limit for the assay was 21 microg mL(-1) (50 nmol mL(-1)) for 5 microL injected (R.S.D. 6.4%, n = 8). A 5 cm short Chiralcel OD column was used. Determination of enantiomeric purity of the drug in aqueous samples required increased sensitivity. The sample was acidified and extracted into a small volume of 1-pentanol, out of which 25 microl was analyzed by SFC. The minor enantiomer at the 3% (w/w) level added could be confirmed. Its ratio remained constant during the procedure as measured relative to a reference solution in organic media.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15296403     DOI: 10.1016/j.chroma.2004.05.035

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  1 in total

1.  Signal enhancement in supercritical fluid chromatography-diode-array detection with multiple injection.

Authors:  Mingzhe Sun; Charlotta Turner; Margareta Sandahl
Journal:  J Sep Sci       Date:  2019-11-19       Impact factor: 3.645