Abstract

Disassembly of bipolar cups with a self-centering system occurred in six patients (seven hips; five women, one man) and the cause of the failure was evaluated. The mean age of the patients at the time of arthroplasty was 49.7 years (range, 27-85 years), and mean weight was 48.4 kg (range, 37-65 kg). The mean time to failure was 7.5 years (range, 4.8-9.2 years). Before failure, all implants functioned well and none of the patients had sustained trauma. Retrieval study showed that the cause of failure of the locking mechanism with a self-centering system was severe polyethylene abrasion at the rim attributable to impingement and deformity of the locking ring. After the locking ring detached, the inner head dislocated from the outer head resulting in a varus position of the outer head in the acetabulum. If the deformed locking ring did not detach, the inner head could dislocate, with the varus outer head remaining in the acetabulum. The incidence of this failure was 11%. This disassembly is not a rare occurrence numerous years after a well-functioning bipolar hemiarthroplasty, even with the self-centering system.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15292802     DOI: 10.1097/00003086-200408000-00022

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  4 in total

1.  What Is the Frequency of Early Dissociation of Bipolar Cups and What Factors Are Associated With Dissociation?

Authors:  Young-Kyun Lee; Chan Ho Park; Yong-Chan Ha; Kyung-Hoi Koo
Journal:  Clin Orthop Relat Res       Date:  2018-08       Impact factor: 4.176

2.  Treatment of femoral neck fractures in elderly patients over 60 years of age - which is the ideal modality of primary joint replacement?

Authors:  Christian Ossendorf; Max J Scheyerer; Guido A Wanner; Hans-Peter Simmen; Clément Ml Werner
Journal:  Patient Saf Surg       Date:  2010-10-20

3.  Five cases of early dissociation between the bipolar hip endoprosthesis cup components; either spontaneously or during reduction maneuvers.

Authors:  Vedat Uruç; Raif Özden; İbrahim Gökhan Duman; Aydıner Kalacı
Journal:  Acta Orthop Traumatol Turc       Date:  2017-02-21       Impact factor: 1.511

4.  Disassembly of Components of a Monoblock Bipolar Hip Prosthesis Following Dislocation: A Case Report and Review on "Bottle Opener Effect".

Authors:  Mukesh Kumar Saini; Neelam Ramana Reddy; Pera Jayavardhan Reddy
Journal:  J Orthop Case Rep       Date:  2020-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.