Variability of left ventricular ejection fraction and volumes with quantitative gated SPECT: influence of algorithm, pixel size and reconstruction parameters in small and normal-sized hearts.

Anne-Sophie Hambye1, Ann Vervaet, André Dobbeleir.   

Abstract

PURPOSE: Several software packages are commercially available for quantification of left ventricular ejection fraction (LVEF) and volumes from myocardial gated single-photon emission computed tomography (SPECT), all of which display a high reproducibility. However, their accuracy has been questioned in patients with a small heart. This study aimed to evaluate the performances of different software and the influence of modifications in acquisition or reconstruction parameters on LVEF and volume measurements, depending on the heart size.
METHODS: In 31 patients referred for gated SPECT, 64(2) and 128(2) matrix acquisitions were consecutively obtained. After reconstruction by filtered back-projection (Butterworth, 0.4, 0.5 or 0.6 cycles/cm cut-off, order 6), LVEF and volumes were computed with different software [three versions of Quantitative Gated SPECT (QGS), the Emory Cardiac Toolbox (ECT) and the Stanford University (SU-Segami) Medical School algorithm] and processing workstations. Depending upon their end-systolic volume (ESV), patients were classified into two groups: group I (ESV>30 ml, n=14) and group II (ESV<30 ml, n=17). Agreement between the different software packages and the influence of matrix size and sharpness of the filter on LVEF and volumes were evaluated in both groups.
RESULTS: In group I, the correlation coefficients between the different methods ranged from 0.82 to 0.94 except for SU-Segami (r=0.77), and were slightly lower for volumes than for LVEF. Mean differences between the methods were not significant, except for ECT, with which LVEF values were systematically higher by more than 10%. Changes in matrix size had no significant influence on LVEF or volumes. On the other hand, a sharper filter was associated with significantly larger volume values though this did not usually result in significant changes in LVEF. In group II, many patients had an LVEF in the higher range. The correlation coefficients between the different methods ranged between 0.80 and 0.96 except for SU-Segami (r=0.49), and were slightly worse for volumes than for LVEF values. In contrast to group I, however, inter-method variability was quite large and most mean LVEF differences were significant. LVEF was systematically highest with ECT and lowest with SU-Segami. With QGS, changes in matrix size from 64(2) to 128(2) were associated with significantly larger volumes as well as lower LVEF values. Increasing the filter cut-off frequency had the same effect. With SU-Segami, a larger matrix was associated with larger end-diastolic volumes and smaller ESVs, resulting in a highly significant increase in LVEF. Increasing the filter sharpness, on the other hand, had no influence on LVEF though the measured volumes were significantly larger.
CONCLUSION: In patients with a normal-sized heart, LVEF and volume estimates computed from different commercially available software packages for quantitative gated SPECT are well correlated. LVEF and volumes are only slightly sensitive to changes in matrix size. Smoothing, by contrast, is associated with significant changes in volumes but usually not in LVEF values. However, owing to the specific characteristics of each algorithm, software should not be interchanged for follow-up in an individual patient. In small hearts, on the other hand, both the used software and the matrix size or smoothing significantly influence the results of quantitative gated SPECT. LVEF values in the higher range are frequently observed with all the studied software except for SU-Segami. A larger matrix or a sharper filter could be suggested to enhance the accuracy of most commercial software, more particularly in patients with a small heart.

Entities:  

Mesh:

Year:  2004        PMID: 15290114     DOI: 10.1007/s00259-004-1601-2

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  19 in total

1.  Gated SPET quantification of small hearts: mathematical simulation and clinical application.

Authors:  K Nakajima; J Taki; T Higuchi; M Kawano; M Taniguchi; K Maruhashi; S Sakazume; N Tonami
Journal:  Eur J Nucl Med       Date:  2000-09

2.  An investigation of the estimation of ejection fractions and cardiac volumes by a quantitative gated SPECT software package in simulated gated SPECT images.

Authors:  A D Achtert; M A King; S T Dahlberg; P H Pretorius; K J LaCroix; B M Tsui
Journal:  J Nucl Cardiol       Date:  1998 Mar-Apr       Impact factor: 5.952

3.  Reproducibility and accuracy of gated SPECT for determination of left ventricular volumes and ejection fraction: experimental validation using MRI.

Authors:  E Vallejo; D P Dione; W L Bruni; R T Constable; P P Borek; J P Soares; J G Carr; S G Condos; F J Wackers; A J Sinusas
Journal:  J Nucl Med       Date:  2000-05       Impact factor: 10.057

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Modelling the integration of myocardial regional perfusion and function.

Authors:  M L Goris; C Thompson; L J Malone; P R Franken
Journal:  Nucl Med Commun       Date:  1994-01       Impact factor: 1.690

6.  Evaluation of left ventricular function and volumes in patients with ischaemic cardiomyopathy: gated single-photon emission computed tomography versus two-dimensional echocardiography.

Authors:  E C Vourvouri; D Poldermans; J J Bax; G Sianos; F B Sozzi; A F Schinkel; J de Sutter; G Parcharidis; R Valkema; J R Roelandt
Journal:  Eur J Nucl Med       Date:  2001-11

7.  Accuracy of ventricular volume and ejection fraction measured by gated myocardial SPECT: comparison of 4 software programs.

Authors:  K Nakajima; T Higuchi; J Taki; M Kawano; N Tonami
Journal:  J Nucl Med       Date:  2001-10       Impact factor: 10.057

8.  Comparison of automatic quantification software for the measurement of ventricular volume and ejection fraction in gated myocardial perfusion SPECT.

Authors:  D P Lum; M N Coel
Journal:  Nucl Med Commun       Date:  2003-03       Impact factor: 1.690

9.  Repeatability of automatic left ventricular cavity volume measurements from myocardial perfusion SPECT.

Authors:  G Germano; P B Kavanagh; J T Kavanagh; S H Wishner; D S Berman; G J Kavanagh
Journal:  J Nucl Cardiol       Date:  1998 Sep-Oct       Impact factor: 5.952

10.  Impact of Wiener filter in determining the left ventricular volume and ejection fraction using thallium-201 gated SPECT.

Authors:  A Manrique; A Hitzel; I Gardin; J N Dacher; P Vera
Journal:  Nucl Med Commun       Date:  2003-08       Impact factor: 1.690

View more
  20 in total

1.  Software-dependent processing variability in SPECT functional parameters: Clinical implications.

Authors:  Saurabh Malhotra; Prem Soman
Journal:  J Nucl Cardiol       Date:  2016-02-17       Impact factor: 5.952

2.  Evaluation of left ventricular volumes and ejection fraction by gated SPECT and cardiac MRI in patients with dilated cardiomyopathy.

Authors:  Feng Wang; Jian Zhang; Wei Fang; Shi-Hua Zhao; Min-Jie Lu; Zuo-Xiang He
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-04-18       Impact factor: 9.236

3.  Gated blood-pool SPECT versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction.

Authors:  François Harel; Vincent Finnerty; Jean Grégoire; Bernard Thibault; François Marcotte; Patricia Ugolini; Paul Khairy
Journal:  J Nucl Cardiol       Date:  2010-02-12       Impact factor: 5.952

4.  Detection of infarct size safety threshold for left ventricular ejection fraction impairment in acute myocardial infarction successfully treated with primary percutaneous coronary intervention.

Authors:  Roberto Sciagrà; Fabrizio Cipollini; Valentina Berti; Angela Migliorini; David Antoniucci; Alberto Pupi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-01-10       Impact factor: 9.236

5.  Normal limits of ejection fraction and volumes determined by gated SPECT in clinically normal patients without cardiac events: a study based on the J-ACCESS database.

Authors:  Kenichi Nakajima; Hideo Kusuoka; Shigeyuki Nishimura; Akira Yamashina; Tsunehiko Nishimura
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-01-12       Impact factor: 9.236

6.  Reducing the small-heart effect in pediatric gated myocardial perfusion single-photon emission computed tomography.

Authors:  Hiroto Yoneyama; Kenichi Nakajima; Koichi Okuda; Shinro Matsuo; Masahisa Onoguchi; Seigo Kinuya; Lars Edenbrandt
Journal:  J Nucl Cardiol       Date:  2016-05-19       Impact factor: 5.952

7.  Ready, shoot, aim? Summary justice for small hearts in nuclear cardiology.

Authors:  Guido Germano; Paul B Kavanagh
Journal:  J Nucl Cardiol       Date:  2016-05-20       Impact factor: 5.952

8.  Development of a new Python-based cardiac phantom for myocardial SPECT imaging.

Authors:  Osama S Hanafy; Magdy M Khalil; Ibrahim M Khater; Haitham S Mohammed
Journal:  Ann Nucl Med       Date:  2020-10-17       Impact factor: 2.668

9.  Transient ischemic dilation for coronary artery disease in quantitative analysis of same-day sestamibi myocardial perfusion SPECT.

Authors:  Yuan Xu; Reza Arsanjani; Morgan Clond; Mark Hyun; Mark Lemley; Mathews Fish; Guido Germano; Daniel S Berman; Piotr J Slomka
Journal:  J Nucl Cardiol       Date:  2012-03-08       Impact factor: 5.952

Review 10.  Normal values for nuclear cardiology: Japanese databases for myocardial perfusion, fatty acid and sympathetic imaging and left ventricular function.

Authors:  Kenichi Nakajima
Journal:  Ann Nucl Med       Date:  2010-01-29       Impact factor: 2.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.