Alessia Panusa1, Monica Ottaviani, Mauro Picardo, Emanuela Camera, Luigi Gagliardi, Paola Chimenti, Arianna Granese, Domenica Tonelli.
Abstract
A high performance liquid chromatography electrospray mass spectrometry (HPLC-ESI-MS) method, for the detection of corticosteroids in cosmetics has been developed. A water-acetonitrile linear gradient on a C-18 reversed-phase column was found to be suitable in separating triamcinolone and its main derivatives, which greatly differ in lipophilicity. Detection was performed in negative electrospray ionisation mode. Good correlation between peaks areas and solutions concentration was found in the range 0.05-10.0 micro g ml(-1) and the detection limits resulted in the range of 20-45 pg injected. The method was successfully applied to the analysis of real samples of shampoo.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15284915 DOI: 10.1039/b402905d
Source DB: PubMed Journal: Analyst ISSN: 0003-2654 Impact factor: 4.616