Estimation of errors in determining intrathoracic blood volume using thermal dilution in pigs with acute lung injury and haemorrhage.

M Nirmalan1, M Niranjan, T Willard, J D Edwards, R A Little, P M Dark.   

Abstract

BACKGROUND: Global end diastolic volume (GEDV) has a constant and predictable relationship to intrathoracic blood volume (ITBV). The present study assesses the difference between ITBV derived from GEDV and ITBV measured directly in pigs with acute lung injury (ALI) and mild haemorrhage.
METHODS: We caused ALI in 12 anaesthetized pigs by i.v. injection of oleic acid and removed 10% of estimated blood volume. EVLW, GEDV, ITBV (COLD; Pulsion Medical Systems), Pa(o(2))/Fi(o(2)), lung compliance and haemodynamic variables were measured at baseline (time 0) and at 30 and 120 min. All animals were volume-resuscitated, followed by measurements at 180 min. A linear equation estimated from the 44 pairs of ITBV and GEDV values in 11 animals was applied iteratively to the four GEDV measurements in the 12th animal, enabling 48 comparisons between measured (ITBVm) and derived ITBV (ITBVd) to be made.
RESULTS: Increase in extravascular lung water index (EVLWi) was associated with significant pulmonary hypertension, worsening of oxygenation and compliance (repeated measures ANOVA; P<0.05). There was good within-subject correlation and agreement between ITBV(m) and ITBV(d) (r=0.72, mean bias 0.8 ml; sd 32 ml). Mean error in deriving ITBV from GEDV was 4.5%. (sd 4.2%; range 0.05-19%). There were no significant differences in errors in the presence of small (up to 10%) deficits in blood volume (F=1.0; P=0.41).
CONCLUSIONS: ITBV estimated by thermodilution alone is comparable to measurements made by the thermo-dye dilution technique in the presence of pulmonary hypertension and mild deficits in total blood volume.

Entities:  

Mesh:

Year:  2004        PMID: 15277298     DOI: 10.1093/bja/aeh232

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  4 in total

Review 1.  Measurement of extravascular lung water using the single indicator method in patients: research and potential clinical value.

Authors:  Lisa M Brown; Kathleen D Liu; Michael A Matthay
Journal:  Am J Physiol Lung Cell Mol Physiol       Date:  2009-07-17       Impact factor: 5.464

2.  Resuscitation of hemorrhagic shock using normal saline does not damage the glycocalyx in the immediate resuscitation phase.

Authors:  Antonius Hocky Pudjiadi; Agus Firmansyah; D V M Gunanti Soeyono; Saptawati Bardosono; Sri Widia Jusman; Minarma Siagian; Munar Lubis
Journal:  Turk J Emerg Med       Date:  2022-01-20

3.  Comparison of three methods of extravascular lung water volume measurement in patients after cardiac surgery.

Authors:  Benjamin Maddison; Christopher Wolff; George Findlay; Peter Radermacher; Charles Hinds; Rupert M Pearse
Journal:  Crit Care       Date:  2009-07-06       Impact factor: 9.097

4.  The effects of differential injection sites of cold saline on transpulmonary thermodilution parameter values.

Authors:  Wanjie Yang; Qingguo Feng; Youzhong An; Xuefeng Zhao; Kai Wei; Chang Li; Wei Wang; Hongyun Teng
Journal:  Patient Prefer Adherence       Date:  2015-04-08       Impact factor: 2.711

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.