Abstract

PURPOSE: Thermoradiotherapy has been shown in several randomized trials to increase local control compared to radiotherapy alone. The first randomized study of interstitial hyperthermia in glioblastoma multiforme showed a survival benefit for hyperthermia, though small. Improvement of the heating technique could lead to improved results. The purpose of this feasibility study is to present the clinical and thermal data of application of an improved interstitial hyperthermia system. METHODS AND MATERIALS: Six patients with a glioblastoma multiforme were treated with interstitial hyperthermia using the Multi Electrode Current Source Interstitial Hyperthermia (MECS-IHT) system. The MECS-IHT system has the capability of spatial monitoring of temperature and individually steering of heating electrodes. Three sessions were given aiming at a steady state temperature of 42 degrees C for 1 h, with an interval of 3-4 days, during an external irradiation scheme of 60 Gy in 6 weeks. Hyperthermia was delivered with a mean of 10 catheters, 18 heating electrodes and 38 thermal probes per patient.
RESULTS: Sub-optimal temperatures were encountered in the first two patients leading to adjustments in technique thereafter with subsequent improvement of thermal data. With a catheter spacing of 11-12 mm, measurements yielded a mean T(90), T(50) and T(10) of 39.9, 43.7 and 45.2 degrees C, respectively, over three sessions in the last patient. The power per electrode to reach this temperature distribution varied from 25-100% of full power in each of the last four patients. Thermal data were reproducible over the three sessions. Acute toxicity was minimal.
CONCLUSIONS: Despite the spatial steering capabilities of the MECS-IHT system, a large temperature heterogeneity was encountered. The heterogeneity was the reason to limit the catheter spacing to 11-12 mm, thus making only small tumour volumes feasible for interstitial heating.

Entities:  

Mesh:

Year:  2004        PMID: 15277019     DOI: 10.1080/02656730410001668357

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  5 in total

1.  Intracranial thermotherapy using magnetic nanoparticles combined with external beam radiotherapy: results of a feasibility study on patients with glioblastoma multiforme.

Authors:  Klaus Maier-Hauff; Ronny Rothe; Regina Scholz; Uwe Gneveckow; Peter Wust; Burghard Thiesen; Annelie Feussner; Andreas von Deimling; Norbert Waldoefner; Roland Felix; Andreas Jordan
Journal:  J Neurooncol       Date:  2006-06-14       Impact factor: 4.130

Review 2.  Heating technology for malignant tumors: a review.

Authors:  H Petra Kok; Erik N K Cressman; Wim Ceelen; Christopher L Brace; Robert Ivkov; Holger Grüll; Gail Ter Haar; Peter Wust; Johannes Crezee
Journal:  Int J Hyperthermia       Date:  2020       Impact factor: 3.914

Review 3.  Integrating Hyperthermia into Modern Radiation Oncology: What Evidence Is Necessary?

Authors:  Jan C Peeken; Peter Vaupel; Stephanie E Combs
Journal:  Front Oncol       Date:  2017-06-30       Impact factor: 6.244

4.  Hyperthermia Treatment Planning Including Convective Flow in Cerebrospinal Fluid for Brain Tumour Hyperthermia Treatment Using a Novel Dedicated Paediatric Brain Applicator.

Authors:  Gerben Schooneveldt; Hana Dobšíček Trefná; Mikael Persson; Theo M de Reijke; Klas Blomgren; H Petra Kok; Hans Crezee
Journal:  Cancers (Basel)       Date:  2019-08-15       Impact factor: 6.639

5.  Experimental iodine-125 seed irradiation of intracerebral brain tumors in nude mice.

Authors:  Joost J C Verhoeff; Lukas J A Stalpers; Annet W Coumou; Kees Koedooder; Cristina Lavini; Cornelis J F Van Noorden; Jaap Haveman; William P Vandertop; Wouter R van Furth
Journal:  Radiat Oncol       Date:  2007-09-26       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.