Allograft impaction and cement penetration after revision hip replacement. A histomorphometric analysis in the cadaver femur.

H Frei1, P Mitchell, B A Masri, C P Duncan, T R Oxland.   

Abstract

We studied various aspects of graft impaction and penetration of cement in an experimental model. Cancellous bone was removed proximally and local diaphyseal lytic defects were simulated in six human cadaver femora. After impaction grafting the specimens were sectioned and prepared for histomorphometric analysis. The porosity of the graft was lowest in Gruen zone 4 (52%) and highest in Gruen zone 1 (76%). At the levels of Gruen zones 6 and 2 the entire cross-section was almost filled with cement. Cement sometimes reached the endosteal surface in other Gruen zones. The mean peak impaction forces exerted with the impactors were negatively correlated with the porosity of the graft.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15274278     DOI: 10.1302/0301-620x.86b5.14016

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  4 in total

1.  Factors influencing the stability of stems fixed with impaction graft in vitro.

Authors:  Hirotsugu Ohashi; Masanori Matsuura; Tsuneyuki Ebara; Yusaku Okamoto; Hironori Kou
Journal:  Clin Orthop Relat Res       Date:  2009-01-29       Impact factor: 4.176

Review 2.  [Influence of thermodisinfection on impaction of cancellous bone : An in vitro model of femoral impaction bone grafting].

Authors:  C Fölsch; A Jahnke; A Groß; G Martels; G A Krombach; M Rickert; M Kampschulte
Journal:  Orthopade       Date:  2018-01       Impact factor: 1.087

3.  Influence of antibiotic pellets on pore size and shear stress resistance of impacted native and thermodisinfected cancellous bone: An in vitro femoral impaction bone grafting model.

Authors:  C Fölsch; J Bok; G A Krombach; M Rickert; C A Fonseca Ulloa; G A Ahmed; M Kampschulte; A Jahnke
Journal:  J Orthop       Date:  2020-09-19

4.  Effect of synthetic bone replacement material of different size on shear stress resistance within impacted native and thermodisinfected cancellous bone: an in vitro femoral impaction bone grafting model.

Authors:  C Fölsch; P Sahm; C A Fonseca Ulloa; G A Krombach; M Kampschulte; M Rickert; A Pruss; A Jahnke
Journal:  Cell Tissue Bank       Date:  2021-04-24       Impact factor: 1.522