A six-bank multi-leaf system for high precision shaping of large fields.

R Topolnjak1, U A van der Heide, B W Raaymakers, A N T J Kotte, J Welleweerd, J J W Lagendijk.   

Abstract

In this study, we present the design for an alternative MLC system that allows high precision shaping of large fields. The MLC system consists of three layers of two opposing leaf banks. The layers are rotated 60 degrees relative to each other. The leaves in each bank have a standard width of 1 cm projected at the isocentre. Because of the symmetry of the collimator set-up it is expected that collimator rotation will not be required, thus simplifying the construction considerably. A 3D ray tracing computer program was developed in order to simulate the fluence profile for a given collimator and used to optimize the design and investigate its performance. The simulations show that a six-bank collimator will afford field shaping of fields of about 40 cm diameter with a precision comparable to that of existing mini MLCs with a leaf width of 4 mm.

Entities:  

Mesh:

Year:  2004        PMID: 15272679     DOI: 10.1088/0031-9155/49/12/012

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Design and fabrication of the control part of a prototype multileaf collimator system.

Authors:  Abdolreza Hashemian; Mohammad Taghi Bahreyni Toossi; Shahrokh Nasseri
Journal:  J Med Signals Sens       Date:  2014-10

2.  Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems.

Authors:  Sanaz Hariri; Majid Shahriari
Journal:  J Appl Clin Med Phys       Date:  2010-06-15       Impact factor: 2.102