New aspects on the role of blood pressure and arterial stiffness in mechanical assistance by intra-aortic balloon pump: in-vitro data and their application in clinical practice.

Theodoros G Papaioannou1, Dimitrios S Mathioulakis, Kimon S Stamatelopoulos, Elias J Gialafos, John P Lekakis, John Nanas, Stamatios F Stamatelopoulos, Sokratis G Tsangaris.   

Abstract

Despite the well-known beneficial effects of the intra-aortic balloon pump (IABP) generally, there are still some clinical conditions accompanied by IABP ineffectiveness. The aim of this study was the investigation of the independent effects of arterial stiffness and blood pressure on acute IABP effectiveness. For this purpose, a mock circulatory system and 20 patients with cardiogenic shock due to acute myocardial infarction, were employed. It was shown that IABP acute efficiency was determined primarily by arterial compliance (AC) rather than blood pressure alone. IABP induced low hemodynamic effects in patients with systolic blood pressure > 80 mm Hg but with increased AC, whereas IABP resulted in greater hemodynamic effectiveness in cases with systolic pressure < 70 mm Hg but lower AC. The present study provides evidence concerning the hemodynamic conditions, which might lead to optimization of IABP or to the prediction of its acute hemodynamic performance, based on both measurements of AC and blood pressure.

Entities:  

Mesh:

Year:  2004        PMID: 15270953     DOI: 10.1111/j.1525-1594.2004.00080.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  4 in total

1.  Intra-aortic balloon pumping reduces the increased arterial load caused by acute cardiac depression, modifying central and peripheral load determinants in a time- and flow-related way.

Authors:  Daniel Bia; Edmundo I Cabrera-Fischer; Yanina Zócalo; Ricardo L Armentano
Journal:  Heart Vessels       Date:  2011-11-09       Impact factor: 2.037

2.  End-diastolic flow reversal limits the efficacy of pediatric intra-aortic balloon pump counterpulsation.

Authors:  Carlo R Bartoli; Benjamin D Rogers; Constantine E Ionan; George M Pantalos
Journal:  J Thorac Cardiovasc Surg       Date:  2013-10-15       Impact factor: 5.209

3.  Discerning aortic waves during intra-aortic balloon pumping and their relation to benefits of counterpulsation in humans.

Authors:  Christina Kolyva; George M Pantalos; Guruprasad A Giridharan; John R Pepper; Ashraf W Khir
Journal:  J Appl Physiol (1985)       Date:  2009-09-03

4.  Measurements of Intra-Aortic Balloon Wall Movement During Inflation and Deflation: Effects of Angulation.

Authors:  Gianpaolo Bruti; Christina Kolyva; John R Pepper; Ashraf W Khir
Journal:  Artif Organs       Date:  2015-05-08       Impact factor: 3.094

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.