Abstract
A new method for density measurements by means of X-ray absorption under high pressure and high temperature using synchrotron radiation has been developed. The method has been modified for a large-volume Paris-Edinburgh press and combined with intense high-energy X-rays at the ESRF. In order to overcome effects of deformation of sample shape under pressure, a ruby cylinder was used as a sample container. The density was determined from the intensity profile of transmitted X-rays. The densities of crystalline and liquid Bi were successfully measured up to 750 K at 1 GPa.Entities:
Year: 1998 PMID: 15263732 DOI: 10.1107/S0909049597015239