Dead-time correction of a multi-element SSD for fluorescent XAFS.

M Nomura1.   

Abstract

The correct evaluation of the live time of a detection system is found to be important for correcting the counting loss of a multi-element detection system for fluorescent XAFS experiments. Synchronous resetting of preamplifiers and the suspension of electronics during the reset period is an effective method. The dead time of the incoming count rate should also be corrected, since its non-linearity cannot be neglected. A 19-element solid-state detector system which can count up to 3.7 x 10(5) counts s(-1) channel(-1) with <270 eV FWHM for Mn Kalpha has been realized; the dead time was independent of the incoming photon energy.

Year:  1998        PMID: 15263674     DOI: 10.1107/S090904959800003X

Source DB:  PubMed          Journal:  J Synchrotron Radiat        ISSN: 0909-0495            Impact factor:   2.616