| Literature DB >> 15261813 |
Dan S T Sit1, Guanghua Gao, Francis C P Law, Paul C H Li.
Abstract
A method was developed for the quantification of matrine in human plasma using a liquid-liquid extraction procedure followed by gas-chromatography-mass spectrometry (GC/MS) analysis. Deuterated matrine, an internal standard of the analysis, was spiked into the plasma samples before extraction. Linear detection responses were obtained for matrine concentrations ranging from 10 to 500 ng/ml. The intra-day and inter-day precision ranged from 0.4 to 4.0% and 1.0-3.5%, respectively. The intra-day accuracy was between -7.3 and 4.5%. The limit of quantification for matrine was 23 ng/ml. The extraction efficiency averaged about 38%. The validated GC/MS method will be used to quantify matrine in human plasma samples collected in a clinical trial study.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15261813 DOI: 10.1016/j.jchromb.2004.05.010
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205