Spin-lattice relaxation and a fast T1-map acquisition method in MRI with transient-state magnetization.

Jung-Jiin Hsu1, Irving J Lowe.   

Abstract

The magnetization under the spin-lattice relaxation and the nuclear magnetic resonance radiofrequency (RF) pulses is calculated for a signal RF pulse train and for a sequence of multiple RF pulse-trains. It is assumed that the transverse magnetization is zero when each RF pulse is applied. The result expressions can be grouped into two terms: a decay term, which is proportional to the initial magnetization M0, and a recovery term, which has no M0 dependence but strongly depends on the spin-lattice relaxation and the equilibrium magnetization Meq. In magnetic resonance pulse sequences using magnetization in transient state, the recovery term produces artifacts and can seriously degrade the function of the preparation sequence for slice selection, contrast weighting, phase encoding, etc. This work shows that the detrimental effect can be removed by signal averaging in an eliminative fashion. A novel fast data acquisition method for constructing the spin-lattice relaxation (T1) map is introduced. The method has two features: (i) By using eliminative averaging, the curve to fit the T1 value is a decay exponential function rather than a recovery one as in conventional techniques; therefore, the measurement of Meq is not required and the result is less susceptible to the accuracy of the inversion RF pulse. (ii) The decay exponential curve is sampled by using a sequence of multiple pulse-trains. An image is reconstructed from each train and represents a sample point of the curve. Hence a single imaging sequence can yield multiple sample points needed for fitting the T1 value in contrast to conventional techniques that require repeating the imaging sequence for various delay values but obtain only one sample point from each repetition. Copyright 2004 Elsevier Inc.

Entities:  

Mesh:

Year:  2004        PMID: 15261622     DOI: 10.1016/j.jmr.2004.05.001

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  1 in total

1.  Characterization of ex vivo healthy human axillary lymph nodes with high resolution 7 Tesla MRI.

Authors:  M A Korteweg; J J M Zwanenburg; P J van Diest; M A A J van den Bosch; P R Luijten; R van Hillegersberg; W P Th M Mali; W B Veldhuis
Journal:  Eur Radiol       Date:  2010-08-09       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.