Literature DB >> 15259711

Intercomparison of monochromatic source facilities for the determination of the relative spectral response of erythemal broadband filter radiometers.

Josef Schreder1, Julian Gröbner, Alexander Los, Mario Blumthaler.   

Abstract

The relative spectral responses of erythemally weighted broadband radiometers determined at three different laboratories are compared, and the systems are described. The results of measurements of four different broadband radiometers are discussed. Although the common dynamic range of the measured relative spectral responses is approximately 10(4), the differences in the relative spectral response functions are lower than 20%. These differences are related mostly to measurement uncertainties and differences in the spectral response facilities.

Year:  2004        PMID: 15259711     DOI: 10.1364/ol.29.001455

Source DB:  PubMed          Journal:  Opt Lett        ISSN: 0146-9592            Impact factor:   3.776


  1 in total

Review 1.  UV Monitoring for Public Health.

Authors:  Mario Blumthaler
Journal:  Int J Environ Res Public Health       Date:  2018-08-11       Impact factor: 3.390

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.