Monte Carlo calculation of the sensitivity of a commercial dose calibrator to gamma and beta radiation.

Jean-Pascal Laedermann1, Jean-François Valley, Shelley Bulling, François O Bochud.   

Abstract

The detection process used in a commercial dose calibrator was modeled using the GEANT 3 Monte Carlo code. Dose calibrator efficiency for gamma and beta emitters, and the response to monoenergetic photons and electrons was calculated. The model shows that beta emitters below 2.5 MeV deposit energy indirectly in the detector through bremsstrahlung produced in the chamber wall or in the source itself. Higher energy beta emitters (E > 2.5 MeV) deposit energy directly in the chamber sensitive volume, and dose calibrator sensitivity increases abruptly for these radionuclides. The Monte Carlo calculations were compared with gamma and beta emitter measurements. The calculations show that the variation in dose calibrator efficiency with measuring conditions (source volume, container diameter, container wall thickness and material, position of the source within the calibrator) is relatively small and can be considered insignificant for routine measurement applications. However, dose calibrator efficiency depends strongly on the inner-wall thickness of the detector.

Mesh:

Substances:

Year:  2004        PMID: 15259666     DOI: 10.1118/1.1755181

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Assessment of the relative contribution of volume and concentration changes in Yttrium-90 labelled resin microspheres on ionization chamber measurements.

Authors:  Nicholas Forwood; Kathy P Willowson; Michael Tapner; Dale L Bailey
Journal:  Australas Phys Eng Sci Med       Date:  2017-11-17       Impact factor: 1.430

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.