Geometric accuracy in robot-assisted total hip replacement surgery.

Jörg Schneider1, Willi Kalender.   

Abstract

OBJECTIVE: Robot-Assisted Surgery in Total Hip Replacement (RAS-THR) has become a fairly widespread procedure for hip replacement surgery in Europe. However, data on the geometric accuracy of the overall process have so far been unavailable. The equipment manufacturers have only published data on the accuracy of certain elements of the procedure (e.g., the accuracy of the robot system). This study aimed to investigate the overall geometric accuracy of the complete process of RAS-THR, as well as the influence of some of the individual component steps.
MATERIALS AND METHODS: An anthropomorphic phantom has been developed representing the hip and upper leg region of a human body. The entire RAS-THR procedure can be carried out using this phantom. Studies were conducted to investigate the overall accuracy of the process, as well as the influence of some individual steps in the procedure, using both the ROBODOC and CASPAR systems.
RESULTS: Overall geometric accuracy of the RAS-THR process was found to be within a standard deviation of 0.5 mm and 0.3 degrees in the critical directions. Most individual steps in the process make similar contributions to the overall variation. The two robot systems examined offer comparable accuracy.
CONCLUSION: Initial results seem to indicate that the RAS-THR process is acceptably accurate. However, acceptance criteria for the necessary geometric accuracy have not yet been defined.

Entities:  

Mesh:

Year:  2003        PMID: 15253366     DOI: 10.3109/10929080309146048

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  3 in total

1.  An experimental evaluation of the force requirements for robotic mastoidectomy.

Authors:  Neal P Dillon; Louis B Kratchman; Mary S Dietrich; Robert F Labadie; Robert J Webster; Thomas J Withrow
Journal:  Otol Neurotol       Date:  2013-09       Impact factor: 2.311

2.  Current concepts in robotic total hip arthroplasty.

Authors:  Pascal Kouyoumdjian; Jad Mansour; Chahine Assi; Jacques Caton; Sebastien Lustig; Remy Coulomb
Journal:  SICOT J       Date:  2020-11-27

Review 3.  Computer-assisted orthopaedic surgery and robotic surgery in total hip arthroplasty.

Authors:  Nobuhiko Sugano
Journal:  Clin Orthop Surg       Date:  2013-02-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.