| Literature DB >> 15250415 |
Masanobu Mori1, Kazuhiko Tanaka, Qun Xu, Mikaru Ikedo, Hiroshi Taoda, Wenzhi Hu.
Abstract
An ion-exclusion chromatography method with ion-exchange enhancement of conductivity was developed for the selective separation and sensitive determination of hydrazine ion from alkali/alkaline earth metal cations and ammonium ion. Hydrazine ion was separated by ion-exclusion/penetration effect from other cations on a weakly basic anion-exchange column in the OH- form (TSKgel DEAE-5PW). Moreover, two different ion-exchange resin columns were inserted between the separating column and conductimetric detector in order to improve the sensitivity of hydrazine ion. The first enhancement column packed with a strongly basic anion-exchange resin in the SO4(2-) form (TSKgel SAX) for hydrazine ion can convert from N2H5OH to (N2H5)2SO4. Moreover, the second enhancement column packed with a strongly acidic cation-change resin in the H+ form (TSKgel SCX) can convert to H2SO4. As a result, the sensitivity of hydrazine ion using two conductivity enhancement columns could be 26.8-times greater than using the separating column alone. This method was effectiveness also for the enhancement of ammonium ion (6.1-times) and sodium ion (1.2-times). The calibration graph of hydrazine ion detected as H2SO4 was linear over the concentration range of 0.001-100 ppm (r2 = 0.9988). The detection limit of hydrazine ion in this system was 0.64 ppb. Therefore, hydrazine ion in real boiler water sample could be accurately determined, avoiding the interference of other cations.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15250415 DOI: 10.1016/j.chroma.2004.03.075
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759