Literature DB >> 15249079

Reliability of tendon excursion measurements in patients using a color Doppler imaging system.

Johannes N M Soeters1, Marij E Roebroeck, Wim P J Holland, Steven E R Hovius, Hendrik J Stam.   

Abstract

PURPOSE: The purpose of this study was to assess the test-retest reliability of tendon excursion measurements with color Doppler imaging in patients with flexor tendon injuries following a modified Kleinert protocol.
METHODS: One observer performed repeated measurements at 3 different time periods in 13 patients with flexor tendon injuries, following a modified Kleinert protocol. The intraclass correlation coefficient (ICC), the standard error of measurement (SEM), and related indices of measurement error were calculated.
RESULTS: Measurements at 10 day after surgery had an ICC of.88 and an SEM of 1.1 mm. Measurements at 6 weeks after surgery had an ICC of.58 and an SEM of 2.0 mm. The measurements after 3 months had an ICC of.94 and an SEM of 1.2 mm.
CONCLUSION: Measurements at 10 days and more than 3 months after surgery were reliable and were as reliable as the measurements performed on healthy subjects. At 6 weeks after surgery the measurements were less reliable. Color Doppler imaging is a reliable and noninvasive method to assess tendon excursion, even in patients with small tendon excursion movements.

Entities:  

Mesh:

Year:  2004        PMID: 15249079     DOI: 10.1016/j.jhsa.2004.04.021

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  4 in total

1.  Ultrasonographic assessment of flexor tendon mobilization: effect of different protocols on tendon excursion.

Authors:  Jan-Wiebe H Korstanje; Johannes N M Soeters; Ton A R Schreuders; Peter C Amadio; Steven E R Hovius; Henk J Stam; Ruud W Selles
Journal:  J Bone Joint Surg Am       Date:  2012-03-07       Impact factor: 5.284

Review 2.  Pathomechanics and Management of Secondary Complications Associated with Tendon Adhesions Following Flexor Tendon Repair in Zone II.

Authors:  Shrikant J Chinchalkar; Juliana Larocerie-Salgado; Nina Suh
Journal:  J Hand Microsurg       Date:  2016-07-21

3.  Reliability of ultrasound speckle tracking with singular value decomposition for quantifying displacement in the carpal tunnel.

Authors:  Verena J M M Schrier; Stefanie Evers; Johan G Bosch; Ruud W Selles; Peter C Amadio
Journal:  J Biomech       Date:  2019-01-21       Impact factor: 2.712

4.  Vibration imaging for localization of functional compartments of the extrinsic flexor muscles of the hand.

Authors:  Yogesh K Mariappan; Armando Manduca; Kevin J Glaser; Jun Chen; Kimberly K Amrami; Richard L Ehman
Journal:  J Magn Reson Imaging       Date:  2010-06       Impact factor: 4.813

  4 in total