Dynamic fatigue resistance of implant-abutment junction in an internally notched morse-taper oral implant: influence of abutment design.

Murat Cavit Cehreli1, Kivanç Akça, Haldun Iplikçioğlu, Saime Sahin.   

Abstract

PURPOSE: To compare dynamic fatigue resistance of one- and two-piece abutments connected to internally notched morse-taper oral implants.
MATERIAL AND METHODS: Sixteen synOcta ITI implants were tightened into two metal models. SynOcta abutments for cement-retained restorations and solid abutments were divided into two equal groups and were torque-tightened into the implants. Each implant received a cement-retained crown. Cyclic dynamic axial and lateral peak loads of 75+/-5 N were applied on the implants for a duty of 500,000 cycles at 0.5 Hz, and at an angle of 20 degrees. Prior to the experiments and at each 100,000 cycles of loading, periotest values (PTVs) were measured. Removal torque values (RTVs) of the abutments were also measured with a custom-made torque device at the termination of the experiment.
RESULTS: All abutments and implants were clinically immobile and without any signs of mechanical failure. The final PTVs for both abutments were similar and the difference between groups was insignificant (P>0.05). The RTVs of solid abutments were significantly higher than synOcta abutments (P<0.05).
CONCLUSION: Predictable long-term clinical results can be achieved with solid abutments and synOcta abutments for cement-retained restorations. Solid abutments possess higher removal torque resistance than synOcta abutments when connected to synOcta ITI implants.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15248881     DOI: 10.1111/j.1600-0501.2004.01023.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  5 in total

1.  The change of rotational freedom following different insertion torques in three implant systems with implant driver.

Authors:  Joo-Hyun Kwon; Chong-Hyun Han; Sun-Jai Kim; Jae-Seung Chang
Journal:  J Adv Prosthodont       Date:  2009-03-31       Impact factor: 1.904

2.  Numerical simulation of in vivo intraosseous torsional failure of a hollow-screw oral implant.

Authors:  Murat Cehreli; Murat Akkocaoglu; Kivanc Akca
Journal:  Head Face Med       Date:  2006-11-04       Impact factor: 2.151

3.  Influence of abutment design on retention of metal copings cemented to implants.

Authors:  Albano Porto da Cunha; Glauco Pereira Moysés; Ana Christina Claro Neves; Rafael Pino Vitti; Flávia Cardoso da Rosa Goulart; Laís Regiane da Siva-Concílio
Journal:  Acta Biomater Odontol Scand       Date:  2016-01-26

4.  Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections.

Authors:  Saied Nokar; Habib Hajimiragha; Leyla Sadighpour; Azam Sadat Mostafavi
Journal:  Dent Res J (Isfahan)       Date:  2020-12-10

5.  A three-dimensional finite element study on the stress distribution pattern of two prosthetic abutments for external hexagon implants.

Authors:  Wagner Moreira; Caio Hermann; Jucélio Tomás Pereira; Jean Anacleto Balbinoti; Rodrigo Tiossi
Journal:  Eur J Dent       Date:  2013-10
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.