Literature DB >> 15248877

Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants.

Roland Glauser1, Lars Sennerby, Neil Meredith, Andreas Rée, AnnaKarin Lundgren, Jan Gottlow, Christoph H F Hämmerle.   

Abstract

OBJECTIVES: The objective of this study was to analyze the development of implant stability by repeated resonance frequency analysis (RFA) measurements during 1 year in 23 patients treated according to an immediate/early-loading protocol. The objective was also to evaluate the possible differences between failing and successful implants.
MATERIAL AND METHODS: Eighty-one Brånemark System implants were placed in 23 patients for immediate/early-occlusal loading in all jaw regions. Thirty of the implants were placed in extraction sockets and 62 were subjected to GBR procedures. Apart from clinical and radiographic examinations, the patients were followed with RFA at placement, prosthesis connection and after 1-3, 6 and 12 months. Statistical analyses were carried out to study the possible differences between implants that failed during the study period and implants that remained successful.
RESULTS: Nine implants failed (11.2%) during the 1 year of loading. RFA showed a distinct different pattern between the implants that remained stable and the implants that were lost. The implants that failed during the course of the study showed a significantly lower stability already after 1 month.
CONCLUSION: Within the limitations of this study, it is concluded that failing implants show a continuous decrease of stability until failure. Low RFA levels after 1 and 2 months seem to indicate an increased risk for future failure. This information may be used to avoid implant failure in the future by unloading implants with decreasing degree of stability with time as diagnosed with the RFA technique.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15248877     DOI: 10.1111/j.1600-0501.2004.01036.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  46 in total

1.  The effect of implant shape and bone preparation on primary stability.

Authors:  Sang-Hyun Moon; Heung-Sik Um; Jae-Kwan Lee; Beom-Seok Chang; Min-Ku Lee
Journal:  J Periodontal Implant Sci       Date:  2010-10-31       Impact factor: 2.614

2.  Clinical study on the primary stability of two dental implant systems with resonance frequency analysis.

Authors:  Annette Rabel; Steffen Gerhard Köhler; Andrea Maria Schmidt-Westhausen
Journal:  Clin Oral Investig       Date:  2007-03-31       Impact factor: 3.573

3.  Immediate vs non-immediate loading post-extractive implants: a comparative study of implant stability quotient (ISQ).

Authors:  L Milillo; C Fiandaca; F Giannoulis; L Ottria; A Lucchese; F Silvestre; M Petruzzi
Journal:  Oral Implantol (Rome)       Date:  2016-11-13

4.  Predictive values of resonance frequency analysis as a diagnostic tool in palatal implant loss.

Authors:  Katja Wieczorek; Winfried Harzer; Heinrich Wehrbein; Maximilian Moergel; Martin Kunkel; Britta A Jung
Journal:  Angle Orthod       Date:  2019-03-18       Impact factor: 2.079

5.  Measurements of Implant Stability Following Sinus Lift: A Pilot Clinical Study.

Authors:  Muhamed Ajanović; Alma Kamber-Ćesir; Adis Hamzić; Selma Tosum
Journal:  Acta Stomatol Croat       Date:  2015-09

6.  [Effect of osseointegration in the posterior area by Shenggu pills].

Authors:  Luo Cuifen; Peng Guoguang
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2017-10-01

7.  Evaluation of implant screw loosening by resonance frequency analysis with triaxial piezoelectric pick-up: in vitro model and in vivo animal study.

Authors:  Makiko Miyashita; Toru Ogawa; Hideki Naito; Aya Shibamoto; Angelina Shuhua Wang; Kenta Shobara; Keiichi Sasaki
Journal:  Clin Oral Investig       Date:  2018-04-28       Impact factor: 3.573

8.  Early outcome of an implant system with a resorbable adhesive calcium-phosphate coating--a prospective clinical study in partially dentate patients.

Authors:  Victor Palarie; Constantin Bicer; Karl M Lehmann; Mohammad Zahalka; Florian G Draenert; Peer W Kämmerer
Journal:  Clin Oral Investig       Date:  2011-08-06       Impact factor: 3.573

9.  In vivo evaluation of quantitative percussion diagnostics for determining implant stability.

Authors:  Lindsey R VanSchoiack; Veronica I Shubayev; Robert R Myers; Cherilyn G Sheets; James C Earthman
Journal:  Int J Oral Maxillofac Implants       Date:  2013 Sep-Oct       Impact factor: 2.804

10.  Extension of hydrophilicity stability by reactive plasma treatment and wet storage on TiO2 nanotube surfaces for biomedical implant applications.

Authors:  Marcel F Kunrath; André L M Vargas; Patrícia Sesterheim; Eduardo R Teixeira; Roberto Hubler
Journal:  J R Soc Interface       Date:  2020-09-30       Impact factor: 4.118