Air leakage around endotracheal tube cuffs.

A Dullenkopf1, A Schmitz, M Frei, A C Gerber, M Weiss.   

Abstract

BACKGROUND AND
OBJECTIVE: To compare the recently introduced Microcuff endotracheal tube with conventional tubes in respect of the cuff pressures required to prevent air leakage.
METHODS: The following tubes (ID 7.0mm) were compared: Microcuff HVLP ICU, Mallinckrodt HiLo, Portex Profile Soft Seal, Rüsch Super Safety Clear and Sheridan CF. Fifty patients undergoing endotracheal intubation with a cuffed tube of internal diameter 7.0 mm were studied. Tracheas were intubated using one of the endotracheal tubes in random order. Cuff pressure to prevent air leakage at standardized ventilator setting (peak inspiratory pressure 20 cmH2O/PEEP 5 cmH2O/respiratory rate 15 breaths min(-1)) was assessed by auscultation of audible sounds at the mouth. Patients characteristics and cuff pressures from each brand were compared to the Microcuff group using the Mann-Whitney U-test (P < 0.05 was chosen as the level of statistical significance).
RESULTS: Patients' median age (range) was 14.2 (12.0-17.1) yr, body weight 57.5 (40.0-81.9) kg and length 164.9 (146.5-190.0) cm. No significant differences in patients' characteristics were found between groups. Mean cuff pressure (all tubes) required for air sealing was 19.1 (8-42) cmH2O. The Microcuff tube required significantly lower sealing pressures (9.5 (8-12) cmH2O) compared to the other brands of endotracheal tube (P < 0.05, Mann-Whitney U-test).
CONCLUSION: The Microcuff endotracheal tube with its ultra-thin polyurethane cuff membrane required the lowest sealing pressure to prevent air leakage. These features are potentially of interest for long-term intubated patients and for cuffed endotracheal tubes in children, allowing tracheal sealing at lower cuff pressures implying less damage to the trachea.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15248624     DOI: 10.1017/s0265021504006064

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  12 in total

1.  Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop model.

Authors:  Islem Ouanes; Aissam Lyazidi; Pierre Eric Danin; Nerlep Rana; Annalisa Di Bari; Fekri Abroug; Bruno Louis; Laurent Brochard
Journal:  Intensive Care Med       Date:  2011-02-12       Impact factor: 17.440

Review 2.  Technologic advances in endotracheal tubes for prevention of ventilator-associated pneumonia.

Authors:  Juan F Fernandez; Stephanie M Levine; Marcos I Restrepo
Journal:  Chest       Date:  2012-07       Impact factor: 9.410

3.  Migration of polyurethane high-volume low-pressure cuffed endotracheal tubes after neck flexion and extension.

Authors:  Devanand Mangar; Collin J Sprenker; Rachel A Karlnoski; Robert K Dodson; Bill H Brashears; Katheryne L Downes; Enrico M Camporesi
Journal:  J Anesth       Date:  2013-03-06       Impact factor: 2.078

4.  Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range.

Authors:  Mary Lou Sole; Xiaogang Su; Steve Talbert; Daleen Aragon Penoyer; Samar Kalita; Edgar Jimenez; Jeffery E Ludy; Melody Bennett
Journal:  Am J Crit Care       Date:  2011-03       Impact factor: 2.228

5.  Comparison of the intracuff pressures of three different tracheostomy tubes.

Authors:  Tomoki Nishiyama
Journal:  J Anesth       Date:  2005       Impact factor: 2.078

Review 6.  Cuffed versus uncuffed endotracheal tubes for neonates.

Authors:  Vedanta Dariya; Luca Moresco; Matteo Bruschettini; Luc P Brion
Journal:  Cochrane Database Syst Rev       Date:  2022-01-24

7.  Endotracheal tube cuff pressure assessment maneuver induces drop of expired tidal volume in the postoperative of coronary artery bypass grafting.

Authors:  Douglas W Bolzan; Solange Guizilini; Sonia M Faresin; Antonio Cc Carvalho; Angelo Av De Paola; Walter J Gomes
Journal:  J Cardiothorac Surg       Date:  2012-06-10       Impact factor: 1.637

8.  The effect of tracheal tube size on air leak around the cuffs.

Authors:  Jin-Young Hwang; Sang-Hyun Park; Sung-Hee Han; Seong-Joo Park; Soo-Kyung Park; Jin-Hee Kim
Journal:  Korean J Anesthesiol       Date:  2011-07-21

9.  Changes in endotracheal tube cuff pressure during laparoscopic surgery in head-up or head-down position.

Authors:  Chun-Yu Wu; Yu-Chang Yeh; Ming-Chu Wang; Chia-Hsin Lai; Shou-Zen Fan
Journal:  BMC Anesthesiol       Date:  2014-08-31       Impact factor: 2.217

10.  Tracheal tube cuff inflation guided by pressure volume loop closure associated with lower postoperative cuff-related complications: Prospective, randomized clinical trial.

Authors:  Waleed A Almarakbi; Abdullah M Kaki
Journal:  Saudi J Anaesth       Date:  2014-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.