Literature DB >> 15241145

Ankle arthroplasties generate wear particles similar to knee arthroplasties.

Akio Kobayashi1, Yukihide Minoda, Yoshinori Kadoya, Hirotsugu Ohashi, Kunio Takaoka, Charles L Saltzman.   

Abstract

Second-generation total ankle arthroplasties have encouraging medium-term results, but the wear of the joint materials is of concern. The aim of the current study was to examine and compare the size, shape, and concentration of polyethylene particles in synovial fluid with total ankle arthroplasties and established posterior-stabilized total knee arthroplasties. Synovial fluid was obtained from 15 patients with well-functioning total ankle arthroplasties and 11 patients with posterior-stabilized total knee arthroplasties at least 6 months after surgery. Polyethylene particles were isolated and analyzed using scanning electron microscopy. Particle size (equivalent circle diameter) in ankles was 0.81 +/- 0.09 microm (mean +/- standard error) and in knees was 0.78 +/- 0.08 microm. Particle shape (aspect ratio) in ankles was 1.57 +/- 0.04 and in knees was 2.30 +/- 0.22. The particle concentration was 1.02 +/- 0.43 x 10/mL in ankles and 1.13 +/- 0.56 x 10/mL in knees, and the particle concentration and size in total ankle arthroplasties were similar to those in total knee arthroplasties. Total ankle arthroplasties generated significantly rounder particles than total knee arthroplasties. These data suggest that the long-term result of total ankle arthroplasty should be as good as posterior-stabilized total knee arthroplasties in terms of polyethylene wear and the prevalence of osteolysis.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15241145     DOI: 10.1097/01.blo.0000132182.54217.da

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  4 in total

1.  The Salto total ankle arthroplasty: survivorship and analysis of failures at 7 to 11 years.

Authors:  Michel Bonnin; Fabrice Gaudot; Jean-Raphael Laurent; Scott Ellis; Jean-Alain Colombier; Thierry Judet
Journal:  Clin Orthop Relat Res       Date:  2010-07-01       Impact factor: 4.176

2.  Does Coronal Plane Malalignment of the Tibial Insert in Total Ankle Arthroplasty Alter Distal Foot Bone Mechanics? A Cadaveric Gait Study.

Authors:  Brandt C Buckner; Christina J Stender; Matthew D Baron; Jacob H T Hornbuckle; William R Ledoux; Bruce J Sangeorzan
Journal:  Clin Orthop Relat Res       Date:  2020-07       Impact factor: 4.755

Review 3.  The Utility of Isotropic 3D Magnetic Resonance Imaging in Assessing Painful Total Ankle Replacements.

Authors:  Ji Lin; Carolyn M Sofka; Constantine A Demetracopoulos; Hollis G Potter
Journal:  Foot Ankle Orthop       Date:  2022-04-30

4.  Ballooning osteolysis in 71 failed total ankle arthroplasties.

Authors:  Gurpal Singh; Theresa Reichard; Rita Hameister; Friedemann Awiszus; Katja Schenk; Bernd Feuerstein; Albert Roessner; Christoph Lohmann
Journal:  Acta Orthop       Date:  2016-05-19       Impact factor: 3.717

  4 in total