Quality specifications for biochemical markers of myocardial injury.

M Zaninotto1, M Mion, S Altinier, M Forni, M Plebani.   

Abstract

BACKGROUND: The current approach to the diagnosis and monitoring of myocardial damage, recognizes to biochemical markers, and in particular to troponins, a key role being well demonstrated that all elevated values were associated with a worsened prognosis. In 2001, the IFCC Committee on Standardization of Markers of Cardiac Damage published guidelines addressing the quality specifications for troponin assays in order to guarantee an analytical performance satisfying medical requirements and to standardize the quality of commercial methods. We describe how the application of quality specifications may be useful in daily practice, in order to provide advice to clinicians in the investigations of complex clinical cases of patients suffering from myocardial damage.
MATERIALS AND METHODS: The samples from three patients (cases 1-3) admitted to the hospital with symptoms suggestive of cardiac disease, showing high troponin I (cTnI) values not correlated with clinical condition, were investigated in order to verify the accuracy of the laboratory data. The standard of quality specifications related to assay specificity, imprecision and interferences were evaluated using different platforms for cTnI assays, carrying out imprecision profile and specific studies on more common interferents in immunoassays.
RESULTS: The obtained results allow us to demonstrate two cases of false-positive cTnI values attributable to a macrocomplex between a modified "in vivo" cTnI and immunoglobulin G (case 1) and to a presence of heterophilic antibodies affecting the RxL Dimension procedure (case 3). Instead, the accuracy of data obtained in case 2 was evidenced by the imprecision profile obtained in our laboratory and by the comparison of results between different laboratories using same platform.
CONCLUSIONS: The lack of standardization as well as the wide differences in the development of each assay give rise to major concerns regarding cTnI determinations. The laboratory must therefore check the compliance between the analytical characteristics of the method utilised against recommended quality specifications for a reliable understanding of the frequency of false-positive results as well as other serious analytical errors.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15234637     DOI: 10.1016/j.cccn.2004.02.035

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

Review 1.  On the Effect of Heterophilic Antibodies on Serum Levels of Cardiac Troponins: A Brief Descriptive Review.

Authors:  Aleksey Michailovich Chaulin
Journal:  Life (Basel)       Date:  2022-07-24

Review 2.  Interferences from blood collection tube components on clinical chemistry assays.

Authors:  Raffick A R Bowen; Alan T Remaley
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.