Literature DB >> 15234034

Implanted gold markers for position verification during irradiation of head-and-neck cancers: a feasibility study.

Bram van Asselen1, Homan Dehnad, Cornelis P J Raaijmakers, Jan J W Lagendijk, Chris H J Terhaard.   

Abstract

PURPOSE: To assess the toxicity and reliability of the use of implanted gold markers for position verification during irradiation of head-and-neck cancer. METHODS AND MATERIALS: Ten patients with localized head-and-neck tumors received two gold markers in the parapharyngeal region. The acute and late radiation-related toxicity were scored prospectively using the Common Toxicity Criteria. The patients were immobilized during irradiation using a five-point mask. The marker location was detected in portal images taken with an a-Si flat panel imager. The intermarker distance, as well as the interfraction motion of the markers, was determined for all patients.
RESULTS: No acute major complications were observed. The acute toxicity grade was not greater than normally detected. The markers were visible in all images. On average, the projected intermarker distance varied 0.8 mm (1 standard deviation). A small time trend was observed in the intermarker distance for 3 patients. For these patients, at least one marker was located in the mucosa or pharyngeal constrictor muscle. Deeper-seated gold markers did not show a time trend in the intermarker distance. The random positioning uncertainty determined using the markers was on average 1.1 and 1.4 mm (1 SD) in the craniocaudal and AP direction, respectively.
CONCLUSION: The use of implanted gold markers for position verification during radiotherapy for head-and-neck cancer patients seems safe and feasible. To avoid any chance of migration, markers should be placed in deep muscular compartments.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15234034     DOI: 10.1016/j.ijrobp.2004.01.022

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  2 in total

1.  A simulation technique for computation of the dosimetric effects of setup, organ motion and delineation uncertainties in radiotherapy.

Authors:  Bongile Mzenda; Mir Hosseini-Ashrafi; Antony Palmer; Honghai Liu; David J Brown
Journal:  Med Biol Eng Comput       Date:  2010-04-23       Impact factor: 2.602

2.  Accurate positioning for head and neck cancer patients using 2D and 3D image guidance.

Authors:  Hyejoo Kang; Dale M Lovelock; Ellen D Yorke; Sergey Kriminski; Nancy Lee; Howard I Amols
Journal:  J Appl Clin Med Phys       Date:  2010-10-27       Impact factor: 2.102

  2 in total