Literature DB >> 15227160

Structural failure of a Starr-Edwards aortic track valve.

R E Ringel1, A L Moulton, J E Burns, J I Brenner, M A Berman.   

Abstract

Structural failure of a Model 2400 Starr-Edwards aortic track valve occurred suddenly, 4 years after implantation. At operation, the valve cage was removed from the descending aorta. Examination of the excised prosthesis disclosed minimal cloth wear and no evidence of infective growth; however, three struts were fractured above their insertion into the valve ring. To our knowledge, this type of valve malfunction has not been previously noted.

Entities:  

Year:  1983        PMID: 15227160      PMCID: PMC341611     

Source DB:  PubMed          Journal:  Tex Heart Inst J        ISSN: 0730-2347


  3 in total

1.  Pathologic anatomy of cardiac valve replacement: a study of 224 necropsy patients.

Authors:  W C Roberts; B H Bulkley; A G Morrow
Journal:  Prog Cardiovasc Dis       Date:  1973 May-Jun       Impact factor: 8.194

2.  Late results of valve replacement surgery. II. Complications of prosthetic heart valves.

Authors:  E S Murphy; F E Kloster
Journal:  Mod Concepts Cardiovasc Dis       Date:  1979-11

3.  Early results with composite strut caged ball prostheses.

Authors:  Q Macmanus; G Grunkemeier; L Housman; C Maloney; B Harlan; L Lambert; A Starr
Journal:  Am J Cardiol       Date:  1980-10       Impact factor: 2.778

  3 in total
  1 in total

1.  Asymptomatic strut fracture in DeBakey-Surgitool aortic valves: incidence, management, and metallurgic aspects.

Authors:  J Von Der Emde J; U Eberlein; J Breme J
Journal:  Tex Heart Inst J       Date:  1990
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.