R E Ringel1, A L Moulton, J E Burns, J I Brenner, M A Berman.
Abstract
Structural failure of a Model 2400 Starr-Edwards aortic track valve occurred suddenly, 4 years after implantation. At operation, the valve cage was removed from the descending aorta. Examination of the excised prosthesis disclosed minimal cloth wear and no evidence of infective growth; however, three struts were fractured above their insertion into the valve ring. To our knowledge, this type of valve malfunction has not been previously noted.Entities:
Year: 1983 PMID: 15227160 PMCID: PMC341611
Source DB: PubMed Journal: Tex Heart Inst J ISSN: 0730-2347