Literature DB >> 15224459

Symptomatic vascular complications after vascular closure device use following diagnostic and interventional catheterisation.

C Thalhammer1, M Aschwanden, C Jeanneret, K H Labs, K A Jäger.   

Abstract

BACKGROUND: Haemostatic puncture closure devices for rapid and effective hemostasis after arterial catheterisation are a comfortable alternative to manual compression. Implanting a collagen plug against the vessel wall may become responsible for other kind of vascular injuries i.e. thrombotic or stenotic lesions and peripheral embolisation. The aim of this paper is to report our clinically relevant vascular complications after Angio-Seal and to discuss the results in the light of the current literature. PATIENTS AND METHODS: We report the symptomatic vascular complications in 17 of 7376 patients undergoing diagnostic or therapeutic catheterisation between May 2000 and March 2003 at the University Hospital Basel.
RESULTS: Most patients presented with ischaemic symptoms, arterial stenoses or occlusions and thrombotic lesions (n = 14), whereas pseudoaneurysms were extremely rare (n = 3). Most patients with ischaemic lesions underwent vascular surgery and all patients with a pseudoaneurysm were successfully treated by ultrasound-guided compression.
CONCLUSIONS: Severe vascular complications after Angio-Seal are rare, consistent with the current literature. There may be a shift from pseudoaneurysms to ischaemic lesions.

Entities:  

Mesh:

Year:  2004        PMID: 15224459     DOI: 10.1024/0301-1526.33.2.78

Source DB:  PubMed          Journal:  Vasa        ISSN: 0301-1526            Impact factor:   1.961


  2 in total

1.  Critical lower limb ischemia from an embolized Angio-Seal closure device.

Authors:  Chris Cianci; Robert C Kowal; Georges Feghali; Stephen Hohmann; Robert C Stoler; James W Choi
Journal:  Proc (Bayl Univ Med Cent)       Date:  2013-10

2.  Fluoroscopic guided fogarty embolectomy for an angio-seal embolism in the popliteal artery.

Authors:  Doran Hong; Seung Hwa Lee; Hwan Hoon Chung; Bo Kyoung Seo; Sang Hoon Cha; Kee Yeol Lee; Jeong Cheon Ahn
Journal:  Korean J Radiol       Date:  2013-07-17       Impact factor: 3.500

  2 in total