Literature DB >> 15216189

Clinical experience in 1040 patients with double-velour knitted Dacron vascular prostheses: With particular reference to dilatation and aneurysm formation.

Denton A. Cooley1, Aswath Subram, Dena P. Houchin.   

Abstract

Recent reports of dilatation and aneurysm formation in Dacron fabric grafts have prompted us to review our experience with 1040 patients who received Meadox-Cooley double-velour knitted grafts over a 47-month period. Bifurcation grafts were used in 398 patients with aorto-femoral occlusive disease and in 203 patients with aortoiliac occlusive disease. Straight tube grafts were implanted in 310 patients with abdominal aortic aneurysms. Small caliber straight tube grafts were used for femoral-femoral bypass in 112 patients. The remaining 17 patients received double-velour grafts for restoration of the renal (14) and superior mesenteric (3) artery circulation. In a review of patients, no dilatation or aneurysm formation was disclosed by clinical examination, sonography or aortography.

Entities:  

Year:  1981        PMID: 15216189      PMCID: PMC287947     

Source DB:  PubMed          Journal:  Cardiovasc Dis        ISSN: 0093-3546


  16 in total

1.  Aneurysm occurring in a femoral artery Dacron prosthesis five and one-half years after insertion.

Authors:  W G KNOX
Journal:  Ann Surg       Date:  1962-11       Impact factor: 12.969

2.  Failure of Dacron prostheses caused by structural defect.

Authors:  J J Yashar; M H Richman; J Dyckman; M Witoszka; R J Burnard; A K Weyman; J Yashar
Journal:  Surgery       Date:  1978-11       Impact factor: 3.982

3.  Arterial grafts: past, present, and future.

Authors:  W S Edwards
Journal:  Arch Surg       Date:  1978-11

4.  True aneurysms in prosthetic femoropopliteal grafts.

Authors:  H Tutassaura; A N Gerein; J G Sladen
Journal:  Am Surg       Date:  1978-05       Impact factor: 0.688

5.  Aneurysm formation in expanded polytetrafluoroethylene prostheses.

Authors:  C D Campbell; D H Brooks; M W Webster; R P Bondi; J C Lloyd; M F Hynes; H T Bahnson
Journal:  Surgery       Date:  1976-05       Impact factor: 3.982

6.  The Sparks' mandril graft. A seven year follow-up of mandril grafts placed by Charles H. Sparks and his associates.

Authors:  R W Hallin; W R Sweetman
Journal:  Am J Surg       Date:  1976-08       Impact factor: 2.565

7.  Failure of ultralightweight knitted Dacron grafts in arterial reconstruction.

Authors:  L W Ottinger; R C Darling; L S Wirthlin; R R Linton
Journal:  Arch Surg       Date:  1976-02

8.  Polytetrafluoroethylene graft aneurysms: a report of five aneurysms.

Authors:  L L Mohr; L L Smith
Journal:  Arch Surg       Date:  1980-12

9.  Postoperative alterations in size of Dacron aortic grafts: an ultrasonic evaluation.

Authors:  D B Nunn; M H Freeman; P C Hudgins
Journal:  Ann Surg       Date:  1979-06       Impact factor: 12.969

10.  Multiple aneurysms in dacron velour graft.

Authors:  J May; M Stephen
Journal:  Arch Surg       Date:  1978-03
View more
  2 in total

1.  Left ventricular aneurysm with 1- to 2-mm-thick myocardium: a variant of the classic true aneurysm?

Authors:  D Liotta; M Del Río; A Gallo; L Frank; A Tamashiro; R Schneider
Journal:  Tex Heart Inst J       Date:  1990

Review 2.  Nonanastomotic aneurysm formation in a Dacron arterial graft: report of a case.

Authors:  I Han; H Shigematsu; M Nunokawa; H Matsuzaki; K Iwata; S Nakamura; M Ishimaru; A Sugiura; Y Kobayashi; Y Morioka
Journal:  Surg Today       Date:  1994       Impact factor: 2.549

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.