| Literature DB >> 15213821 |
Jurg M Daniel1, Sille Ehala, Sebastian D Friess, Renato Zenobi.
Abstract
A new technique is presented for the coupling of atmospheric pressure matrix-assisted laser desorption/ionization (AP-MALDI) mass spectrometry with liquid delivery systems. Mass measurements of polymers and peptides are demonstrated using a co-dissolved matrix, e.g. alpha-cyano-4-hydroxycinnamic acid (HCCA). Improvements in terms of sensitivity are achieved by optimizing the shape und control of the exit capillary and by using a laser (355 nm) at a 1 kHz repetition rate. Two calibration experiments promise a good applicability of the presented coupling method for quantitative measurements. The limit of detection achieved so far is 500 nM for peptides in methanol solution containing 25 mM HCCA.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15213821 DOI: 10.1039/b404178j
Source DB: PubMed Journal: Analyst ISSN: 0003-2654 Impact factor: 4.616