Literature DB >> 15213700

[Creation and clinical application of real-time dose monitor using dose area product meter].

Kosuke Matsubara1, Kichiro Koshida, Atsushi Fukuda, Yoshinori Uoyama, Hiroji Iida, Takashi Mizushima.   

Abstract

The management of patient dose has become more of an issue in recent years. Dose can be determined non-invasively and in real time through the use of a dose area product meter, but it is the area dose value that is obtained. Therefore, we created a program that estimates entrance skin dose (ESD) in real time from area dose values obtained during procedures. We used Microsoft Visual C++ 6.0 (Standard Edition) for the programming language and C language for the programming environment. The value was a maximum 285.4 mGy at ileus tube insertion when measuring ESD for radiography of the digestive organ and non-vascular type IVR using the created program and seeking the average according to the procedures. The program that we created can be considered valid for monitoring ESD correctly and in real time.

Entities:  

Mesh:

Year:  2004        PMID: 15213700     DOI: 10.6009/jjrt.kj00000922441

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Estimation of the life expectancy of a filament of the conventional X-ray unit: a technical case study.

Authors:  Atsushi Fukuda; Kosuke Matsubara; Tosiaki Miyati
Journal:  Radiol Phys Technol       Date:  2014-10-11
  1 in total