Magdalena Gozdowska1, Ewa Kulczykowska.
Abstract
A new HPLC assay for plasma arginine-vasotocin (AVT) and isotocin (IT) determination based on fluorescence detection preceded by combination of solid-phase extraction (SPE) and fluorescence derivatization is presented. Plasma samples retained on solid support were purified and then derivatized by the fluorescent compound 4-fluoro-7-nitro-2,1,3-benzoxadiazole (NBD-F). The peptide derivatives were eluted from cartridges, pre-concentrated and analyzed by HPLC system with fluorescent detection. The separation was carried out on a reversed-phase column with solvent gradient system. The assay was linear in the range 15-220 pmol ml(-1) for AVT r2=0.998 and 10-220 pmol ml(-1) for IT r2=0.996. The detection limits for AVT and IT were 0.8 and 0.5 pmol ml(-1) (3:1, signal-to-noise), respectively. The recoveries of derivatized hormones were in the range 89-93%. Both of the inter- and intra-day assay precision were below 5.5 and 9% for AVT and IT, respectively. The assay should be also applicable to plasma and tissue samples from other animals with only minor modification.Entities: